occur in analytical columns. Historically, five detectors have been used. They are the


electron capture detector (ECD), Hall micro-electrolytic conductivity detector (HECD),


thermionic detectors (NPD and AFID), and the flame photometric detector (FPD).


ECD measures the loss of detector electrical current produced by a sample component


containing electron-absorbing molecule(s). This detector is very sensitive for measuring


halogenated pesticides, in the analysis of chlorinated hydrocarbon pesticides


(organochlorines) such as aldrin, dieldrin and DDT. ECD is efficient for the analysis


of poly chlorinated biphenyls (PCBs) as well. The HECD can measure chlorine (and


other halogens), nitrogen, or sulphur. This detector is more selective than the ECD,


though the ECD is more sensitive. The Hall electrolytic conductivity detector also


has improved over the last few years, and has replaced the ECD in those laboratories


where extreme sensitivity is not required. Both the NPD and AFID measure the


presence of nitrogen and phosphorus atoms in the pesticide, with little response


resulting from other types of atoms in the molecules.


Today, the flame photometric detector (FPD) measures sulphur or phosphorus,


and is a rugged, highly stable, and very selective detector, since it does not detect


compounds other than organophosphates and those containing sulphur. The flame


photometric detector is less sensitive for phosphorus than the NPD and less sensitive


for sulphur than the Hall detector. However, it is useful for the analysis of unclean


crude herbal extracts. Conventional mass spectrometers (MS) have also been used by


some pesticide residue laboratories as gas chromatography detectors, and as high-


performance liquid chromatography detectors as well. MS is normally used when


special techniques are necessary to confirm the identity of a particular pesticide,


when conventional detectors cannot detect the pesticide. The use of MS is growing,


especially with the development of the more portable and less costly mass selective


detector (MSD). The MSD and ion trap detector (ITD) may become more routinely


used for pesticide residue analysis, as improvements in their computer software are


made and their scan parameters become more suitable for chromatography.


High-performance liquid chromatography (HPLC) for the analysis of pesticide


residues is a fairly recent technology, but it is becoming the second most frequently


used technique after GC. GC depends upon the volatilisation of the pesticide, whereas


HPLC is dependent on the stationary phases that can selectively retain any molecular


structure; polar, non-polar, ionic, or neutral. Separations can even occur as a function


of molecular size (gel permeation) or chemical derivatisations (synthesis of a chemical


derivative of the pesticide). HPLC is not as efficient as capillary gas chromatography


for separator purposes because the chromatographic peaks are broader, though HPLC


columns are more efficient than packed GC columns when columns of equal length


are considered. HPLC columns usually last longer because they are not subjected to


the extremely high temperatures that GC columns are. The HPLC detectors used for


pesticide residue analysis are the UV absorption, fluorometer, conductivity, and


electrochemical.


Many pesticides absorb UV light at the wavelength of mercury discharge (254


nanometres) and can be detected in very small quantities. Unfortunately, many food


co-extractives do so as well, making this detector nearly useless for trace analysis in


foods. An alternative is the variable wavelength detector, which can be tuned to a


wavelength that is absorbed by the pesticide but not by the food co-extractives. The


fluorometer is a highly sensitive HPLC detector for some pesticides, which is typically


used for pesticides with aromatic molecular structures such as alachlor or paraquat.


This detector, however, has limited application to the detection of most pesticides