STACK SAMPLING 1105
the probe liner is regulated with a variable voltage controller.
The controller is calibrated by introducing air at various tem-
peratures in the range of the stack temperature and deter-
mining the controller setting necessary to maintain the probe
temperature required by the Method (120 14ºC) or the
specific regulation, such as a subpart of 40 CFR Part 60. The
probe liner, with the heating element, is inserted in a stain-
less steel tube.
The probe liner is cleaned in much the same way as the
nozzle. Once it has cooled it is removed from the train and the
ends wiped free of silicone grease and covered. During clean-
ing, the probe is loosely held on an incline with utility clamps.
The probe is then rinsed with acetone while slowly rotating
the probe. A nylon brush, extended with teflon tubing, is used
to scrub the inside of the probe liner. The acetone is saved in
the bottle labeled Container 2 for later analysis.
Pitot Tube
A type S pitot tube is secured to the outside of the probe.
The pitot tube was discussed in the Method 2 description.
The special relationship to the nozzle is critical, in terms of
both the location and the alignment. The pitot tube should
be inspected before and after each run to insure that it has
not been damaged. The pitot tube and the tubing connecting
it to the manometer should be leak checked before and after
each run. This can easily be done by slipping rubber tubing
over the end of one side of the pitot, blowing gently on the
rubber tubing producing at least 3 inches water pressure, then
clamping the rubber tubing with a pinch clamp. No change
in pressure should be observed in 15 seconds.
Filter
The filter holder is made of borosilicate glass with a glass
frit filter support and silicone rubber gasket. Clamps of vary-
ing designs are used to seal the filter between the two halves
of the holder.
The filter is a glass fiber filter capable of capturing at
least 99.95% of 0.3 micron particles. Filters, desiccated at
room temperature and ambient pressure, are weighted every
six hours until the change is less than 0.5 mg. This process
is called desiccating to constant weight. Alternatively the
filters can be dried in an oven to constant weight. They are
allowed to cool to room temperature in a desiccator before
each weighing.
After the test, the filters are inspected for signs of gases
passing around the edges of the filter and for water stains
on the filter. Either would seriously compromise the results.
The filter is carefully transferred to a petri dish. Any pieces
of filter sticking to the housing are removed with forceps or a
sharp instrument. The front half of the filter holder is rinsed
with acetone and the acetone saved in the bottle labeled
Container 2 for later analysis.
The filter holder must be maintained at 120 14ºC, or
as specified in the applicable regulation.
Any heating system capable of providing this tempera-
ture around the filter holder during the run is acceptable.
TEMPERATURE SENSOR
TEMPERATURE
SENSOR
PROBE
PROBE
PITOT TUBE
PITOT MANOMETER
TYPE S
PITOT TUBE
ORIFICE
THERMOMETERS
DRY GAS METER AIR TIGHT
PUMP
MAIN VALVE
BY-PASS VALVE
IMPINGERS ICE BATH
VACUUM
GAUGE
HEATED AREA
STACK
WALL
THERMOMETER THERMOMETER
FILTER HOLDER
VACUUM
LINE
CHECK
VALVE
IMPINGER TRAIN OPTIONAL, MAY BE REPLACED
BY AN EQUIVALENT CONDENSER
Schematic of Method 5 sampling train.
FIGURE 6 Schematic of Method 5 sampling train.
C019_003_r03.indd 1105C019_003_r03.indd 1105 11/18/2005 11:07:15 AM11/18/2005 11:07:15 AM