23.1 Emission/Absorption Spectroscopy and Energy Levels 955
410.2 (violet)434.0 (violet)486.1 (green-blue) 656.3 (red)
400 500 600 700
/nm
Figure 23.2 The Visible Portion of the Hydrogen Atom Emission Spectrum
(Simulated).Each wavelength represented produces an image of the slit of the spec-
trograph. If only discrete wavelengths are present, as in this case, the spectrum is called a
line spectrum.
Toroid
mirror
Toroid
mirror
Paraboloid
mirror
Dual grating
(on turntable)
Thermocouple
Filter Exit slit
Ellipsoid
mirror
Entrance slit
Plane mirror
Plane
mirrors
Optical wedge
reference beam
Sampling area
Sample beam
100% comb
Plane mirror
Toroid
mirror
Aperture
stop
Sector
mirror
Plane
mirrors
Source
Figure 23.3 Schematic Diagram of a Filter-Grating, Double-Beam Infrared Spec-
trophotometer.This diagram shows a scanning instrument. The wavelength passed is
determined by the angular position of the grating, which is mounted on a turntable, and is
rotated automatically by a motor. Courtesy of Perkin-Elmer Corporation.
Absorption Spectroscopy
Absorption spectroscopy has traditionally been carried out in aspectrophotometer,
such as the one shown schematically in Figure 23.3. The light is dispersed by a prism
or grating, collimated into a beam of nearly parallel rays, and passed through a cell
containing the sample. Only a narrow band of wavelengths passes at one time and the
bandwidth determines the resolution of the instrument. The wavelength can be chosen
by turning the prism or grating, and this is done automatically in a scanning instrument.