ACGTACGT ACG
Direction of electrophoretic movement
T
23
Fig. 5.38
Autoradiograph of a DNA sequencing gel. Samples were prepared using the Sanger dideoxy method of DNA sequencing. Each set of four samples was loaded in
to adjacent
tracks, indicated by A,C, G and T, depending on the identity of the dideoxyribonucleotide used for that sample. Two sets of samples were labelled with
35 S (1 and 3) and one
was labelled with
32 P (2). It is evident that
32 P generates darker but more diffuse bands than does
35 S, making the bands nearer the bottom of the autoradiograph easy to see.
However, the broad bands produced by
32 P cannot be resolved near the top of the autoradiograph, making it impossible to read a sequence from this region. The much sharper bands
produced by
35 S allow sequences to be read with confidence along most of the autoradiograph and so a longer sequence of DNA can be obtained from a single gel.