9780521516358book.pdf

(lily) #1
Quadrupole rods

Ion that is unsuccessful
at reaching detector

Detector

RF supply

DC supply

+

+







Ions

Fig. 9.5Quadrupole analyser. The fixed (DC) and oscillating (RF) fields cause the ions to undergo complicated
trajectories through the quadrupole filter. For a given set of fields, only certain trajectories are stable, which
only allows ions of specificm/zto travel through to the detector. The efficiency of the quadrupole is
impaired after a build-up of ions that do not reach the detector. Therefore a set of pre-filters is added to the
quadrupole to remove the ions that would otherwise affect the main quadrupole.

100

Relative ion intensity (%)


900

0
1000 1100
m/z

1200 1300 1400 1500

%

A100
1027.6

A106
988.0

A106
969.4

A97
1059.4A94
1093.1

A86
1154.6

A86
1194.6

A80
1284.0
A78
1317.3
A76
1352.1

M= 102658.3 ±7.0

Fig. 9.6Large intact protein mass accurately measured in electrospray MS. The species of ions are
annotated by the charge state, e.g. with 99, 100, 101 charges, etc., and the associatedm/zvalue. The inset
shows the ‘deconvoluted spectrum’.

361 9.3 Mass analysers