Illustrated Guide to Home Chemistry Experiments

(Amelia) #1
Chapter 5: Mastering Laboratory Skills 79

CoSERN vING TITRANT
If you are certain that the titration will require much
less than the full capacity of the burette, you can conserve
your titrating solution by partially filling the burette. Make
sure that you fill it far enough to allow the tip to be filled
and air bubbles to be eliminated. For example, if you
are using a 50 mL burette and have calculated that the
titration should require about 20 mL of titrant, fill the
burette to about 30 mL and run a few mL through the tip
to eliminate air bubbles. Leave yourself some slack. If you
run out of titrant before you’ve neutralized the analyte,
you’ll either have to repeat the titration from scratch
or add more titrant to the burette, which reduces the
accuracy of your titration.

waste solutions) and remove the burette from the clamp. Rinse the
burette thoroughly inside and out with tap water, making sure to
run the rinse water through the tip. If the burette uses a pinchcock,
remove the pinchcock and rubber tubing. If it uses a ground
glass stopcock, remove the stopcock. If it uses a Teflon stopcock,
release the tension on the stopcock and/or remove the stopcock,
depending on the design of the burette. Rinse the burette, tip, and
stopcock assemblies thoroughly with tap water and then with
distilled water and place them aside to dry thoroughly.


PRFORMIE nG TITRATIOnS WITHOUT
A BURETTE
If you don’t have a burette, you can use one of the methods
described in the following sections to perform titrations with
differing levels of convenience and accuracy.


TITRATE USING A GRAdUATEd CyLINdER
If your only volumetric glassware is a graduated cylinder, you can
use it to perform quick-and-dirty titrations with fair accuracy. To
titrate using your graduated cylinder, take the following steps:



  1. If possible, calculate the approximate amount of titrant
    you expect to be required to complete the titration.

  2. Fill the graduated cylinder with titrant until the titrant
    reaches the top graduation mark as closely as possible.
    Use a Beral pipette to add the final few drops to bring
    the level of the meniscus as close as possible to the line.

  3. Use the Beral pipette to withdraw about 2 mL of titrant,
    and set the pipette aside. Invert the pipette to make
    sure that none of the titrant is lost.

  4. Slowly trickle titrant from the graduated cylinder into
    the titration vessel with swirling until you have nearly
    reached the equivalence point. You’re approaching the
    equivalence point when the indicator changes color
    where the titrant is being added, but that color change
    disappears as you swirl the titration vessel.


FIGURE 5-10: Reading the meniscus on a burette


AddNGI pARTIAL dRopS
As you near the end point of a titration, you can add
partial drops by manipulating the stopcock carefully. As
you release each partial drop, use a wash bottle filled with
distilled water to flush the partial drop from the tip of the
burette directly into the solution being titrated.

pRELINARmI y TITRATIoNS
If you have no idea where the end point of a titration will
occur, you can save a lot of time by doing a fast trial run
instead of slowly adding titrant from the beginning of the
titration. To do so, run titrant quickly into the titration
vessel until the indicator begins to change color. If you’re
lucky, the titration will not have reached its end point, and
you can begin adding titrant more slowly until you reach
the end point. If you overshoot the mark while adding titrant
quickly, note the reading on the burette and use something
less than that as the starting point for the real titration.
For example, if adding 22.0 mL of titrant was just a bit too
much, when you start the real titration you can quickly run
20.0 mL of titrant into the titrating vessel and then add
titrant drop-by-drop until you reach the end point.

5.hen you are near the equivalence point, stop adding W
titrant from the graduated cylinder, and begin adding
titrant drop-by-drop from the Beral pipette. Continue
adding titrant dropwise from the Beral pipette until you
reach the equivalence point, which is indicated by a
color change that persists for at least 10 or 15 seconds
as you continue to swirl the titration vessel.
Free download pdf