Wine Chemistry and Biochemistry

(Steven Felgate) #1

12 Automatic Analysers in Oenology 663


Characteristics of the Method


Intra-laboratory reproducibility: 4.5
Inter-laboratory reproducibility: 8.5

12.3.4.8 Determination of Iron in Wines and Musts


Principle


Chemical Method


The method is based on the Aquaquant©Rkit for iron commercialised by Merck. Fer-


rous ions react with the reagent, Ferrospectral (the sodium salt off (pyridyl-2)-3 bis


(phenyl-4-sulfonique)-5,6 triazine-1,2,4,acid) forming a violet complex. Only the


ferrous form of iron reacts with this reagent, so the inclusion of an additional reagent


containing ammonium thioglycolate digests weak iron complexes and reduces triva-


lent iron to the bivalent form.


Characteristics of the Method


Intra-laboratory reproducibility: 0.5mg/L


12.3.4.9 Determination ofα-Amino Nitrogen in Wines and Musts


Principle


Chemical Method


To t a lα-amino nitrogen is determined by its reaction witho-phthaldialdehyde. The


intensity of the absorbance 340 nm is compared to that obtained for standards of


known concentrations of isoleucine.


12.3.4.10 Determination of Ammoniacal Nitrogen in Wines and Musts


Principle


Enzymatic Method


In the presence of reduced nicotinamide-adenosine-dinucleotide (NADH), NH



  • 4
    ions are reacted with 2-oxoglutarate resulting in the formation ofL-glutamate, in


a reaction catalysed by glutamate deshydrogenase (GLDH):


NH
+
4 +2-oxoglutarate+NADH−> L-glutamate+NAD

++H
2 O
( in the presence of GLDH)

The quantity of NADH oxidised into NAD+is proportional to the quantity of


ammoniacal nitrogen present in the sample.