Handbook for Sound Engineers

(Wang) #1
Acoustical Noise Control 73

measured TL curve shown in Fig. 4-10, the STC
overlay is first aligned to 500 Hz and adjusted vertically
to read some estimated value, say, STC-44. The differ-
ence between the measured TL level and the STC curve
is recorded at each of the octave points. These data
are added together. The total, 47 dB, is more than the
32 dB allowed. The STC overlay is next lowered to an
estimated STC-42, and a total of 37 dB results.
Lowering the overlay to STC-41 yields a total of 29 dB,
which fixes the STC-41 contour as the rating for the TL
curve of Fig. 4-10.

The final illustration of STC methods is given in
Fig. 4-11. In this case, a pronounced coincidence dip
appears at 2500 Hz. This illustrates the second STC
requirement, “the maximum deficiency at any single test
point shall not exceed 8 dB.” This 8 dB requirement
fixes the overlay at STC-39, although it might have been
considerably higher if only the 32 dB sum requirement
applied.
The shape of the standard STC contour may be very
different from the measured TL curve. For precise

work, using measured, or even expertly estimated, TL
curves may be desirable rather than relying on STC
single number ratings. Convenience usually dictates use
of the STC shorthand system, but it is, at best, a rather
crude approximation to the real-world TL curves.

Assume that a goal of NC-20 has been chosen for a
sound room. The noise survey indicates a noise level
and spectrum as shown in Fig. 4-12. What wall
construction will bring the noise of Fig. 4-12 down to
the NC-20 goal we have set for the interior? Fig. 4-13
shows that a wall having a rating of STC-55 is required.
The next step is to explore the multitude of possible
wall configurations to meet the STC-55 requirement as
well as other needs.

If the NC curve in Fig. 4-12 is subtracted from the
measured noise curve, this will indicate the raw data
that indicates the amount of loss needed to achieve the
desired NC. This is plotted in Fig. 4-13. The standard
STC template is laid over the graph and the needed STC
is read opposite the 500 Hz mark.

Table 4-5. Standard STC Contour


Frequency
in Hz

Octave Sound
Transmission Loss
in dB

Frequency
(Hertz)

Octave Sound
Transmission Loss
in dB

125 24 800 42
160 27 1000 43
200 30 1250 44
250 33 1600 44
315 36 2000 44
400 39 2500 44
500 40 3150 44
630 41 4000 44

Figure 4-10. The method of determining the single-number
STC rating of a barrier from its measured TL graph.

(^1) » 3
(^1) » 3 1 » 3
50
40
30
(^20) 125 250 500 1k 2k 4k
160 315 630 1.25k 2.5k
200 400 800 1.6k 3.15k
(^1) / 3 octave band center frequency—Hz
3.03.85.25.2 5.0
3.7
2.0 0.8
STC-41
Sound transmission loss—dB
Measured TL
Figure 4-11. The second rule for STC determination that a
maximum deficiency of 8 dB is allowed.
Figure 4-12. Noise spectrum from noise survey.
STC-39
50
40
30
(^20) 125 250 500 1k 2k 4k
160 315 630 1.25k 2.5k
200 400 800 1.6k 3.15k
(^1) / 3 octave band center frequency—Hz
Sound transmission loss—dB
Measured TL
8 dB
32 63 100 200 400 8001.25k 2k 3.1 5k 8k
60
55
50
45
40
30
25
20
15
10
Frequency–Hz
SPL–dB

Free download pdf