1000 Chapter 26
26.2.2 Reference Levels
In the early days of broadcasting and recording, both
10 mW and 12.5 mW into a 500: line were used as a
reference level. However, later this was changed to
6 mW. In May 1939 the present standard of 1 mW into a
600 : line was adopted. This reference level was
selected as a level that would conform to the telephone
company's standards of limiting the signal level on a
transmission line to a value that would produce a
minimum of crosstalk and still provide a satisfactory
signal-to-noise ratio (SNR). The 1 mW reference level
is a unit quantity and is readily applicable to the decimal
system, being related to the watt by the factor 10^3.
Zero level is a reference power level of 1 mW of
power into a 600: load. This is equivalent to a voltage
of 0.775 V.
26.2.3 VU Meter Impedance
The VU meter and its attenuator impress a 7500:
impedance onto a circuit. The VU meter system consists
of an indicator movement, a variable attenuator, and a
series resistor of 3600:, Fig. 26-5. Meter manufac-
turers supply only the meter movement; the external
circuitry is added later. A 200 μA D’Arsonval meter
movement with an internal resistance of 3900: and a
full-wave, copper-oxide or selenium rectifier are
contained within the meter case. The attenuator is vari-
able in steps of 2 dB, presents a constant impedance of
3900 : to the meter movement, and prevents the ballis-
tics of the meter from being affected when the attenu-
ator setting is changed.
Standard VU meters are designed to read 0 VU, or
100%, with 1.228 V (+4 dBm) applied to the instru-
ment. If the meter is used with the attenuator but
without the 3600: series resistor and is connected
across a 600: load in which 1 mW of power is
flowing, the movement will be deflected to the 100%
calibration point. This method is not recommended
since the impedance looking back into the meter is only
3900 : and loads the 600: circuit. It is the usual prac-
tice to keep the impedance of bridging devices at a ratio
of 10:1 or greater.
Increasing the input impedance of the VU meter
from 3900:to 7500: creates a 4 dB loss across the
3600 : resistor. If a signal of 1 mW (0.775 V) is
impressed across the input terminals of the circuit in
Fig. 26-6, it will not deflect the meter to the 0 VU cali-
bration but only to the 4 VU (or decibel) mark, or
approximately 65%. This means that if the meter is to
be deflected to the 100% point, the input signal must be
increased to a +4 dBm. This is the reason why 1 mW of
power will be indicated at the 4 dB calibration mark.
Attenuators used with VU meters start at a +4 dBm.
The bridging loss caused by the VU meter being
inserted into the circuit is the drop in signal level caused
by the absorption of power by the meter circuit. As a
rule, the power absorbed is quite small and may be
ignored. However, at high powers, it may become
important. Bridging loss may be calculated by the
equation
(26-2)
where,
BR is the VU meter input impedance,
Z is the line impedance.
A 7500: VU meter has a bridging loss of 0.34 dB.
26.2.4 VU Impedance Level Correction
VU meters are calibrated for 1 mW of power across a
600 : load as 4 VU, therefore when a VU meter is
connected across any other impedance, a correction
must be added to the indicated reading to give a proper
VU reading. The equation for the level correction is
(26-3)
where,
dBcorr is the decibel amount added to the VU reading,
Z 2 is the impedance for which the meter is calibrated,
Z 1 is the impedance of the circuit bridged.
A typical example of applying a correction factor is
as follows: a VU meter calibrated for a line impedance
of 600: is bridged across a 16: loudspeaker line and
indicates a level of +1 dBm. The true VU would be
.(26-4)
The correction factor from Eq. 26-2 is
Figure 26-5. A 7500 : VU meter, calibrated for 1 mW
reference level or 0.775 V across 600 :.
3600 7
600 7
0.775 V 7
R 1 R 1
R 2
3900 7
Rectifier
meter
3900 7
internal
R (^3) resistance
3900 7 attenuator
dBloss 20
2 BR+Z
2 BR
= log-------------------
dBcorr 10
Z 2
Z 1
= log-----
VU 1 dBm+= correction factor