quantitative manner, care must be taken in its interpretation, as it may be
indicative but not definitive. Some samples, e.g.polished metals, may yield
quantitative results whilst powdery samples will be indicative of relative
concentration at best. The sensitivity of this technique is typically in the
0.1–0.5% range, depending upon element and matrix.
WDS information is collected with a specific crystal detector that is rotated
through the range where each specific element’s diffraction wavelength is
found. Different crystals are also used for different wavelength ranges and the
geometry of the primary beam, sample and crystal are important for proper
detection and collection of the information. WDS is much slower than EDS,
but is generally one order of magnitude more sensitive than EDS. As a result,
both techniques can nicely complement each other, as they have different