Cardiovascular Fitness Procedures Manual

(coco) #1

2.3.2 Calibration of the Automatic Blood Pressure Monitor


The calibration of the pressure amplifier inside the STBP monitor will be checked each week

with a mercury manometer. The reading on the STBP monitor should be within +/- 5mmHg from the


reading on the mercury manometer. If the difference between the readings differ by more than 5 mm Hg,


the STBP monitor will need to be returned to the manufacturer for calibration.


„ Weekly calibration check procedures are outlined below:


  • Turn off the power switch for the STBP monitor.

  • Get portable mercury manometer, and Y-connector with rubber bulb from the
    room supply box.

  • Set up the mercury manometer on the computer cart.

  • Connect one end of the Y-connector to the air tubing on the mercury
    manometer.

  • Connect second end of the Y-connector to the inflation outlet at the back of the
    STBP monitor.


„ Use the following steps to check the calibration of the STBP with the mercury
manometer. The calibration will be checked at these five ranges:


  1. 220-2 50

  2. 180-2 20

  3. 120-1 80

  4. 70-12 0

  5. 0-70


„ While pressing the END switch, turn POWER switch on.

„ Press MANUAL switch to set unit in the MANUAL mode. The word “MAN” is
displayed on the LED of the ELAPSED TIME indicator.

„ While pressing the UP switch, press the START switch.

„ Raise the pressure in the mercury manometer to 250 mmHg. Tighten the screw to
maintain the mercury at a constant level.

„ In about 1 second, the pressure values are displayed on the STBP monitor.
Free download pdf