Instant Notes: Analytical Chemistry

(Tina Meador) #1
melted when the temperatures of the liquid indium and alumina again rise at
the same rate.
Two alternative strategies can now be adopted. If the temperatures of sample
S and reference R are measured and the temperature difference recorded (the
differential thermal analysis or DTAstrategy),

DT =TS−TR


a downward peak (i.e. a minimum) is recorded. Under carefully controlled
instrumental conditions, this may be related to the enthalpy change for the
thermal event:

DHK


f
i

DTdtKA

where Ais the area of the temperature-time peak from initial (i) to final (f)
point.
This leads to quantitative or heat-flux differential scanning calorimetry
(heat-flux DSC). The negative sign is required since the enthalpy change on
melting is positive, but DTfor melting is negative.
The second strategy is to control the amount of heat supplied to sample and
reference so that their temperatures stay as nearly the same as possible. Using
separate heaters for sample and reference allows measurement of the difference
in power DPto be measured. With proper control and calibration, this will give
the enthalpy change of the peak directly:

DH =DPdt


This is known as power-compensated DSC. Essentially, it has been demon-
strated that both strategies produce equivalent results with similar accuracy.

Instrumentation The schematic diagram of the apparatus for DSC/DTA is shown in Figure 1. The
temperature, both for the sample and the reference and also the furnace is
measured by thermocouples, or resistance sensors. Higher sensitivity and
greater stability are obtained if multiple sensors of inert material are used.


312 Section G – Thermal methods


Fig. 1. Schematic diagram of DTA or DSC apparatus (Drepresents either the temperature
or power difference).

Atmosphere
control

Sensor
amplifier

Furnace
Sample
Sensor

Programmer
or computer

Cooling
control

Data recorder
or computer

SR

Free download pdf