Farm Animal Metabolism and Nutrition

(Tina Sui) #1

those things which commonly cause prob-
lems is shown in Table 9.1.
While any of these factors can vary
from one sample to the next, the problem
arises when they vary between the samples
used to develop a calibration, and new
samples on which the calibration is used.
Ultimately, the objective in developing a
calibration is to find a set of samples whose
spectra/composition are representative of a
larger set of samples, i.e. one does not want
any systematic differences. Thus, one does
not want to develop a calibration using
samples ground to one mesh size or
scanned at one temperature, and then use
that calibration to determine the values for
samples ground to a different mesh (or even
ground on a different type of grinder) or
scanned at a different temperature. While it
may not be possible to take into account the
repair of an instrument, one can avoid the
problems caused by the other factors by
scanning the calibration samples over a
range of temperatures or scanning each
sample at several different temperatures,
and so forth. Thus, to a large degree,
variations likely to influence spectra can be


built into the calibration set. The calibra-
tion process will then avoid those wave-
lengths where, for example, variations due
to temperature have too great an influence.
Also, spectral pre-treatments, such as
multiplicative scatter correction (Isaksson
and Kolwalski, 1993), can be used to
eliminate some effects, such as those due to
particle size differences. The real problems
occur with differences in spectrometers and
samples. Spectrometer to spectrometer
variation is difficult and expensive to build
into a calibration because it requires
running the same samples on many
different instruments and using all the
spectra obtained to develop the calibration.
Even then, another spectrometer may vary
too greatly from those used for the calibra-
tion for it to work properly. Therefore,
generally, the problem is handled outside
the calibration development process in
what is called ‘Calibration Transfer’.
Sample differences lie at the heart of
calibration development, and revolve
around ‘Sample Populations’. In developing
calibrations, many effects (temperature
variations, etc.) can be eliminated or

190 J.B. Reeves III


Absorbance

Data points

6 5 4 3 2 1

E

F

D

C

A B

Fig. 9.4.Hypothetical effects of peak shift (B), baseline shifts (C and D) and a second component (E and F)
on an NIR spectrum (A).

Free download pdf