Farm Animal Metabolism and Nutrition

(Tina Sui) #1

reduced by deliberately including the varia-
tion in the calibration; the difficult problem
lies in the nature of the samples themselves
and the variations that occur over a growing
season, between seasons, between growing
locations, between varieties, etc.


Closed and open populations

Closed populations and
calibration development
A closed population is one in which one
has all the samples of interest, e.g. 2000
lucerne samples from a pasture feeding
experiment. After grinding the samples, the
samples should be scanned in a random
order to avoid any pattern between the
samples and the order of scanning, i.e.
avoid scanning all the highest quality
samples on one day, at one temperature, and
the lowest quality samples on another day,
at a different temperature. The more any
differences due to personnel, room condi-
tions, instrument conditions, etc. are
randomly distributed across all the samples,
the better. The next step is to select the
samples for which the analyte of interest


will be measured by the standard technique,
i.e. NDF, ADF by extraction procedures (Van
Soest, 1994). It is these values which will be
used to determine the wavelengths needed
for the NIR-based determination of the
remaining samples. A smaller set of test
samples is also needed in order to test the
calibration after development. These are
only to determine if the calibration/equation
developed works properly on samples not
in the original calibration set.
The question then is how to find
samples in the set which represent all 2000
samples. Several methods have been used
over the years. First, N samples can be
randomly selected. Second, every nth
sample can be used. If this is done, it is
important to avoid a structured data set:
i.e. a set of 2000 samples consisting of 100
sets of 20 samples, each set in order of
increasing NDF content, and every 20th
sample taken for the calibration set. This
would result in only the samples with the
lowest NDF content being included in the
calibration. Under such circumstances, it is
unlikely that the calibration developed
would work on the test set or very many of
the remaining samples. The third method

Use of Near Infrared Reflectance Spectroscopy 191

Table 9.1.Some factors (as differences between samples) which can alter spectra and influence calibration
accuracy.


Factor Spectral effect Solution


Particle size Baseline shifts, multiplicative Mathematical pre-treatments of spectra
scatter


Temperature Spectral peak shifts Include temperature differences in
calibration


Sample packing Spectra do not represent sample Pack uniformly, scan a larger sample


Sample homogeneity Non-representative sample Grind more finely or scan a larger sample


Constituent out of range Inappropriate for calibration Add a sample with wider range of values


Different feed or samples Inappropriate for calibration Add samples to old or develop new
calibration


Spectrometer repaired or Peak shifts, absorbance changes, Redevelop calibration or use calibration
changes somehow multiplicative effects, etc. transfer


Different spectrometer used Peak shifts, absorbance changes, Redevelop calibration or use calibration
multiplicative effects, etc. transfer


Stray light in instrument Inaccurate determination Fix instrument or use constant lighting at
all times

Free download pdf