Quick Review Cards for Medical Laboratory Science

(avery) #1

Test Performance Specifications


and Verification


Laboratory Operations Review 56

CLIA requires labs to verify manufacturer’s analytical performance claims for # 1–4 below on new methods introduced
after April 24, 2003.


PARAMETER EXPLANATION VERIFICATION


  1. Accuracy

  2. Precision

  3. Reportable range

  4. Reference interval


Lab tests samples of known values (controls, calibrators,
proficiency samples, previously tested patient speci-
mens) to see how close results are to known value.

Lab repeatedly tests same samples (on same day &
different days) to see how close the results are.

Lab tests samples with known values at highest &
lowest levels claimed to be accurate by manufacturer.

If manufacturer’s or published reference ranges are
used, lab must test specimens from normal subjects to
verify ranges. Ranges may need to be adjusted to fit
lab’s patient population.

How close measurement is to true value.

Reproducibility. How close results are when same
sample is tested multiple times. (Note: A procedure
can be precise but not accurate.)

Range of values over which lab can verify accuracy
of test system. Also known as linearity.

Formerly called normal value. Can vary for different
patient populations (age, gender, race). Established
by testing minimum of 120 healthy subjects & de-
termining range in which 95% fall. (Note: 5% of
healthy population falls outside of reference range.)
Labs may use manufacturer’s reference ranges or
published reference ranges, if appropriate for their
patient population.

continued...
Free download pdf