After this wave passes through the optical coupler, the returning waves are
Er¼Er0exp½iðkzrxtÞ ð 10 : 2 Þ
from the reference arm and
Es¼Es0exp½iðkzsxtÞ ð 10 : 3 Þ
from the sample arm. Here zrand zsare the path lengths from the optical coupler to
the reference mirror and to a specific depth in the sample, respectively. The
monochromatic wave amplitudes Er0from the mirror and Es0from the sample are
quite different, because only a small amount of light returns from the sample,
whereas the mirror reflects most of the light incident on it. The total electricfield
Etotalat the detector in Fig.10.3is given by the sum of the two returning waves as
Etotal¼Er0exp½iðkzrxtÞ þEs0exp½iðkzsxtÞ ð 10 : 4 Þ
The intensity IDat the detector of the light beams coming from the two arms of
the interferometer is given by the square of the electricfield amplitude
Cornea
(n = 1.376)
Retina
Incoming light
Lens
(n = 1.406)
Aqueous humor
(n = 1.336) Vitreous humor
(n = 1.336)
A-line scan
Retina
Anterior lensPosterior lens
Cornea
Reflectance peaks
Peak locations
Fig. 10.5 Reflectance peaks for an A-line scan of the eye
296 10 Optical Imaging Procedures