161 10
A simple STEM-in-SEM bright-field detector can be
created as shown in. Fig. 10.18. The grid carrying the thin
specimen is placed over an aperture that serves to stop elec-
trons that have suffered an elastic scattering event in the
specimen. The unscattered beam electrons pass through this
aperture and strike a gold-covered surface below, where they
generate strong SE emission, which is then attracted to the
E–T(positive bias) detector, creating a bright-field image. If
the SEM is also equipped with a TTL detector, the nearly pure
SE 1 image that arises from a thin specimen can be collected
with the TTL detector simultaneously with the bright-field
STEM image collected with the E–T(positive bias) detector,
as shown for particles supported on a lacey-carbon film in
. Fig. 10.19.
10.5.3 Eliminate the Use of SEs Altogether: “Low Loss BSEs“
“Low Loss BSEs“
BSEs are usually considered a low resolution signal because
of the substantial delocalization that results from multiple
elastic scattering of the beam electrons at conventional beam
Simple STEM-in-SEM detector
Short
working
distance, ~ 3 mm
Au surface
Through the lens detector: SE image
E-T detector: “STEM Image”
Electrons
scattered by
particle are
cut off by
aperture
Aperture
C-film and
support grid
Unscattered beam electrons
Scattered
beam
electrons
. Fig. 10.18 Schematic
cross section of a STEM-in-SEM
detector that makes use of the
Everhart–Thornley(positive bias)
detector to form a bright-field
STEM image
Aerosol particles collected on lacey carbon
25 keV cold-FEG-SEM
ab
SEM (TTL) STEM (ET)
500 nm
. Fig. 10.19 Aerosol contamination particles deposited on lacey-carbon film and simultaneously imaged with a TTL detector for SE 1 and the
STEM-in-SEM detector shown in. Fig. 10.18 (Example courtesy John Small, NIST)
10.5 · Achieving High Resolution with Secondary Electrons