218
16
measurements are made on the basis of the same “live-time”
so as to achieve constant dose for quantitative measurements.
The level of activity of the EDS is reported to the user as a
percentage “dead-time”:
Deadtime()%=()ICRO− CR /ICR×100%
(16.5)
Dead-time increases as the beam current increases. The
dead time correction circuit can correct the measurement
time over the full dead-time range to 80 % or higher. (Note
that as a component of a quality measurement system, the
dead-time correction function should be periodically
checked by systematically changing the beam current and
comparing the measured X-ray intensity with predicted.)
However, as the dead-time increases and the arrival rate of
X-rays at the EDS increases, coincidence events become
progressively more prominent. This effect is illustrated in
. Fig. 16.10 for a sequence of spectra from a glass with six
SiK-L+
Ca
K-L,
n
ot Cr
K-L
Al
K+
Ca
K-L,
not
V K-L
2M
gK
2A
lK
2S
iK
SiK+
O K
2O
K not Na
K
Count
s
Coincidence
peaks
Coincidence
peaks
K412 (mass frac)
O0.428
Mg 0.11 7
Al 0.0491
Si 0.21 2
Ca 0.10 9
Fe 0.0774
Energy (keV)
20DT
16DT
12DT
10DT
7DT
14 000
12 000
10 000
8 000
6 000
4 000
2 000
0
0246810
K412
E 0 = 20 keV
7% deadtime
Counts
K412 (mass frac)
O 0.428
Mg 0.117
Al 0.0491
Si 0.212
Ca 0.109
Fe 0.0774
Photon energy (keV)
140 000
120 000
100 000
80 000
60 000
40 000
20 000
0
0246810
. Fig. 16.10 Development of coincidence peaks as a function of dead-
time. NIST Standard Reference Material SRM 470 (Mineral Glasses) K412.
(upper) SDD-EDS spectrum at 7 % dead-time showing the characteristic
peaks for O, Mg, Al, Si, Ca, and Fe. (lower) SDD-EDS spectra recorded over
arrange of dead-times showing in-growth of coincidence peaks. Note
elemental misidentifications that are possible
Chapter 16 · Energy Dispersive X-ray Spectrometry: Physical Principles and User-Selected Parameters