Scanning Electron Microscopy and X-Ray Microanalysis

(coco) #1
306

19


the X-ray lines used for analysis.. Table 19.4 shows the
variation of the path length that can occur if one varies the
initial electron beam energy for Al K-L 3 X-rays in Al from
10 to 30 keV and the take-off angle, ψ = 15° and ψ = 60°.
The variation in PL is larger than a factor of 20, from
0.35 μm at the lowest keV and highest take-off angle to
7.7 μm at the highest keV and lowest take-off angle. Clearly
the analyst’s choices of the initial electron beam energy and
the X-ray take-off angle have a major effect on the path length
and therefore the amount of absorption that occurs.

In summary, using appropriate formulations for X-ray
generation with depth or φ(ρz) curves, the effect of absorp-
tion can be obtained by considering absorption of X-rays
from element i as they leave the sample. The absorption cor-
rection, Ai, can be calculated by taking the ratio of the effect of
absorption for the standard, Ai,std, to X-ray absorption for the
unknown, Ai,unk, for each element, i, in the sample. The effect
of absorption can be minimized by decreasing the path length
of the X-rays in the specimen through careful choice of the
initial beam energy and by selecting, when possible, a high
take-off angle.

X-ray Fluorescence, F


Photoelectric absorption results in the ionization of inner
atomic shells, and those ionizations can also cause the emis-
sion of characteristic X-rays. For fluorescence to occur, an
atom species must be present in the target which has a critical
excitation energy less than the energy of the characteristic
X-rays being absorbed. In such a case, the measured X-ray
intensity from this second element will include both the
direct electron-excited intensity as well as the additional
intensity generated by the fluorescence effect. Generally, the
fluorescence effect can be ignored unless the photon energy is
less than 5 keV greater than the critical excitation energy, Ec.
The significance of the fluorescence correction, Fi, can be
illustrated by considering the binary system Fe-Ni. In this sys-
tem, the Ni K-L 3 characteristic energy at 7.478  keV is greater
than the energy for excitation of Fe K radiation,
Ec = 7.11 keV. Therefore, an additional amount of Fe K-L 3 radi-
ation is produced beyond that due to the direct beam on Fe.

. Figure 19.5 shows the effect of fluorescence in the Fe-Ni sys-
tem at an initial electron beam energy of 30 keV and a take-off
angle, ψ, of 52.5°. Under these conditions, the atomic number
effect, ZFe, and the absorption effect, AFe, for Fe K-L 3 are very
close to 1.0. The measured kFe ratio lies well above the first
approximation straight line relationship. The additional inten-
sity is given by the effect of fluorescence. As an example, for a
10 wt% Fe – 90 wt% Ni alloy, the amount of iron fluorescence
is about 25 %.
The quantitative calculation of the fluorescence effect
requires a knowledge of the depth distribution over which the
characteristic X-rays are absorbed. The φ(ρz) curve of elec-
tron-generated X-rays is the starting point for the fluorescence
calculation, and a new φ(ρz) curve for X-ray- generated X-rays
is determined. The electron-generated X-rays are emitted iso-
tropically. From the X-ray intensity generated in each of the
layers Δ(ρz) of the φ(ρz) distribution, the calculation next
considers the propagation of that radiation over a spherical
volume centered on the depth ρz of that layer, calculating the
absorption based on the radial distance from the starting layer
and determining the contributions of absorption to each layer
(ρz) in the X-ray-induced φ(ρz) distribution. Because of the
longer range of X-rays than electrons in materials, the X-ray-
induced φ(ρz) distribution covers a much greater depth, gen-
erally an order of magnitude or more than the electron-induced
. Table 19.4 Path Length, PL, for Al K-L 3 X-rays in Al


E 0 Take-Off Angle, ψ Rx (μm) Path Length, PL,
(μm)

10 15 0.3 1.16
10 60 0.3 0.35
30 15 2.0 7.7
30 60 2.0 2.3

Eo

EDS
Detector

PL = Z COSEC Ψ

Specimen

PL = t

Ψ

Z

. Fig. 19.15 Schematic diagram showing the X-ray absorption path
length in a thick, flat-polished sample: PL = absorption path length;
ψ = X-ray take-off angle (detector elevation angle above surface)


Chapter 19 · Quantitative Analysis: From k-ratio to Composition
Free download pdf