319 20
20.3.3 Analysis of a Minor Constituent
with Peak Overlap From a Major
Constituent
The problem of accurately recovering peak intensities when
overlaps occur is exacerbated when the concentration ratio of
the elements producing the overlapping peaks is large, for
example, a major constituent (C > 0.1 mass fraction) interfering
with a minor (0.01 ≤ C ≤ 0.1) constituent. The high throughput
(>100 kHz output count rate) of SDD-EDS enables collection
of high count EDS spectra in modest collection time (e.g.,
10 million counts in 100 s). Moreover, the high throughput of
SDD-EDS is achieved with stability in both the peak position
(i.e., calibration) and the peak shape (i.e., resolution) across the
entire input count rate range. In simultaneous WDS-EDS
measurements, this SDD-EDS performance been demon-
strated to the spectrum measurement capabilities necessary for
robust MLLS peak-fitting to achieve accurate measurement of
the interfering peak intensities equal to that of WDS on the
spectroscopically resolved peaks (Ritchie et al. 2012 ).
20.3.4 Ba-Ti Interference in BaTiSi 3 O
BaTiSi 3 O 9 (benitoite) provides an example of severe interfer-
ence between two constituents of identical atomic concentra-
tion but with a mass concentration ratio of Ba/Ti = 2.9—Ti
K-L2,3 (4.510 keV) and Ba L3-M4,5 (4.466 keV)—which are
separated by 44 eV, as shown in. Fig. 20.7. DTSA II analysis
of benitoite with Ti and sanbornite (BaSi 2 O 5 ) as fitting refer-
ences and standards is given in. Table 20.7. Note that in this
analysis, O has been directly analyzed with the k-ratio/matrix
corrections protocol and not by the method of assumed stoi-
chiometry. The analytical results are seen to closely match the
stoichiometry of the ideal mineral formula.
20.3.5 Ba-Ti Interference: Major/Minor Constituent Interference in K
Constituent Interference in K2496
Microanalysis Glass
NIST microanalysis research material K2496 glass contains
these same elements, but with Ba as a major constituent
(C = 0.4299 mass fraction) and Ti as a minor constituent
(C = 0.01799 mass fraction), giving an elemental ratio of Ba/
Ti = 23.9.. Figure 20.8a shows the SDD-EDS spectrum and
residual after peak fitting, and. Table 20.8 contains the
results of the analysis. Despite the severe overlap and the
large elemental ratio, the concentration for Ti is measured
with reasonable accuracy. A reasonable question that the
analyst might ask is, If it was not known that the Ti was pres-
ent, could it be detected?. Figure 20.8b shows the fitting
residual for an analysis protocol in which Ti was not fit. The
peaks for Ti K-L2,3 and Ti K-M 3 are revealed in the residual
spectrum.
20.4 The Need for an Iterative Qualitative
and Quantitative Analysis Strategy
The analysis of NIST glass K2496 demonstrates that rigorous
analysis requires an iterative qualitative analysis–quantita-
tive analysis approach. When analyzing an unknown
material, it is likely that some constituents at the minor and
trace level will not be obvious when the first qualitative anal-
ysis is performed due to peak interference from constituents
at higher concentrations. An alternating qualitative–quanti-
tative analytical strategy is required to discover possibly hid-
den minor and trace constituents. In the initial qualitative
analysis, the EDS spectrum is evaluated to identify the major
and minor elemental constituents whose peaks are readily
identifiable. The k-ratio/matrix correction protocol is then
applied with appropriate choices for elemental peak-fitting
references and for standards, and the “residual” spectrum is
constructed that contains the intensity remaining after the
fitted peaks have been subtracted. If all constituents have
been accounted for, this residual spectrum should only con-
sist of the continuum background and possibly also artifact
peaks such as escape and coincidence peaks. However,
because of the relative poor energy resolution of EDS, the
analyst must perform a second qualitative analysis of the
residual spectrum for the presence of previously unrecog-
nized peaks that are associated with constituents that suffer
interference from the higher intensity peaks. If such peaks
are discovered and assigned to an element(s) not previously
recognized, the quantitative analysis must then be repeated
with this element(s) included in the peak-fitting and
. Table 20.5 Analysis of PbS at E 0 = 10 keV with CuS and PbSe
as fitting references and standards; Integrated spectrum count,
0.1–10 keV = 5,482,000; uncertainties expressed in mass fraction.
Analysis performed with Pb M 5 -N6,7 and S K-L2,3
S Pb
Cav (atom frac) 0.4938 0.5062
Z-correction 1.31 0.983
A-correction 1.028 1.056
F-correction 1 1
σ (7 replicates) 0.000953 0.000953
σRel (%) 0.19 % 0.19 %
RDEV (%) −1.20 % 1.2
C (mass frac, single analysis) 0.1306 0.8651
Counting error, std 0.0001 0.0009
Counting error, unk 0.0003 0.001
A-factor error 0.0002 0.0017
Z-factor error 1.50×10–5 0.0001
Combined errors 0.0004 0.0022
20.4 · The Need for an Iterative Qualitative and Quantitative Analysis Strategy