32020
quantification suite of elements. A third iteration may be
necessary to recover constituents present at the trace level
near the limits of detection.20.4.1 Analysis of a Complex Metal Alloy, IN100
IN100
IN100 is a nickel-based superalloy which produces the
EDS spectrum shown in. Fig. 20.9. In the first qualitative
analysis, characteristic X-ray peaks were identified for
Al K; the Ti K-family; the Cr, Co, and Ni K- and L- fami-
lies; and Mo L-family. Analysis with the k-ratio/matrix
correction protocol using pure elements as peak-fitting
references and as standards gave the results shown in. Table 20.9, with the analytical total slightly below unity.
Close inspection of the residual spectrum in. Fig. 20.9
showed an anomaly at the energy of Ti K-M4,5 (4.931 keV))
which closely corresponds to the energy of V K-L2,3
(4.952 keV) with a separation of 21 eV. When V
was included in the suite of fitted elements, the anomaly
in the residual spectrum was eliminated, as shown in
. Fig. 20.10, and a minor V constituent was recovered in the
. Table 20.6 Analysis of MoS 2 at E 0 = 10 keV with CuS and Mo
as fitting references and standards; integrated spectrum count,
0.1–10 keV = 7,326,000; uncertainties expressed in mass fraction.
Analysis performed with Mo L2,3-M4,5 and S K-L2,3
S Mo
Cav (atom frac) 0.6644 0.3356
Z-correction 1.039 0.884
A-correction 1.083 1.024
F-correction 1 1
σ (7 replicates) 0.0022 0.0022
σRel (%) 0.33 % 0.66 %
RDEV (%) −0.34 % 0.70 %
C (mass frac, single analysis) 0.3972 0.6046
Counting error, std 0.0003 0.0003
Counting error, unk 0.0006 0.0014
A-factor error 0.0006 0.0006
Z-factor error 2.80×10–5 4.40×10–5
Combined errors 0.0008 0.0015a 350 000b300 000250 000200 000150 000100 00050 0000
1.5 1.7 1.9 2.1 2.3 2.5
Photon energy (keV)Counts2.72.9 3.13.3MoS2_10kV20nA11%DT100s
Residual_MoS2_10kV20nAMoS2_10kV20nA11%DT100s
20 000 Residual_MoS2_10kV20nA15 00010 0005 0000
1.5 1.7 1.9 2.1 2.3 2.5
Photon energy (keV)Counts2.72.9 3.13.3E 0 = 10 KeV
MoS 2
Fitting residual. Fig. 20.6 a SDD-EDS spectrum of MoS 2 (red) at E 0 = 10 keV (7,326,000 counts) and residual (blue) after DTSA II analysis using CuS and Mo as
fitting references and standards. b Expanded view
Chapter 20 · Quantitative Analysis: The SEM/EDS Elemental Microanalysis k-ratio Procedure for Bulk Specimens, Step-by-Step