Scanning Electron Microscopy and X-Ray Microanalysis

(coco) #1
334

20


Counts

Photon energy (keV)

Corning Glass A
E 0 = 15 keV
100 μm square
50 μm square
20 μm square
5 μm square
2 μm square
1 μm square
spot

20 000

15 000

5 000

0
2.5 2.7 2.9 3.1 3.3 3.5 3.7 3.9 4.1 4.3

10 000

Counts

Photon energy (keV)

Corning Glass A
E 0 = 15 keV
100 μm square
50 μm square
20 μm square
5 μm square
2 μm square
1 μm square
spot

160 000
140 000
120 000
10 000
80 000
60 000

40 000
20 000

0
0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0

CorningA_1mu_15kV15nA
CorningA_2mu_15kV15nA
CorningA_5mu_15kV15nA
CorningA_20mu_15kV15nA
CorningA_50mu_15kV15nA
CorningA_100mu_15kV15nA
CorningA_spot_15kV15nA

CorningA_1mu_15kV15nA
CorningA_2mu_15kV15nA
CorningA_5mu_15kV15nA
CorningA_20mu_15kV15nA
CorningA_50mu_15kV15nA
CorningA_100mu_15kV15nA
CorningA_spot_15kV15nA

. Fig. 20.16 Corning glass A, showing Na and K migration as a function of dose for scanning beams covering various areas (20 keV, 10 nA)


values for the glass, including the alkali elements Na and K,
whereas the point beam results show reductions in the Na and
K concentrations.. Figure 20.17 shows that the measured
sodium and potassium concentrations increase to reach the
synthesized values as the scanned area dimensions are
increased to cover areas above 20 x 20-μm (nominal magnifi-
cation 5 kX) for the particular dose utilized (15 keV, 1500 nA-s).
Thus, while scanning a large homogeneous area obviously
concedes the spatial resolution capability of electron-excited
X-ray microanalysis, this approach may be the most expedient
technique to control and minimize alkali element migration.
Materials that can serve as useful standards for sodium
include certain crystalline minerals such as albite (NaAlSi 3 O 8 )
in which the sodium is much more stable under electron
bombardment. However, even for albite the use of a station-
ary high intensity point beam may produce significant migra-
tion effects, as shown in. Fig. 20.18 for spectra collected
with a stationary point beam as a function of dose (upper)

and at the same dose with a fixed beam and two different
sizes of scanned areas (lower). Thus, the use of a scanned area
rather than a fixed beam may be necessary when collecting a
standard spectrum, even on a crystalline material.

20.6.2 Materials Subject to Mass Loss


During Electron Bombardment—the


Marshall-Hall Method


Thin Section Analysis


The X-ray microanalysis of biological and polymeric speci-
mens is made difficult, and sometimes impossible, by several
forms of radiation damage that are directly caused by the elec-
tron beam. At the beam energies used in the SEM (0.1–
30  keV), it is possible for the kinetic energy of individual
beam electrons to break and/or rearrange chemical bonds.
The radiation damage can release smaller molecules such as

Chapter 20 · Quantitative Analysis: The SEM/EDS Elemental Microanalysis k-ratio Procedure for Bulk Specimens, Step-by-Step
Free download pdf