354
21
Photon energy (keV)
0.0 2.0 4.0 6.0 8.0 10.0 12.0 14.0 16.0 18.0 20.0
. Fig. 21.14 Sequence of measurements for determining pathological
scattering effects by the “in-hole” method. For the same dose, EDS spectra
are measured on the brass block, aluminum stub, platinum aperture, and
at the center of the 200 μm diameter hole of the platinum aperture. A
logarithmic display reveals the very low counts in the “in-hole” spectrum
compared to the spectra measured on the various materials
EDS
detector
window
Collimator &
electron trap
Final
lens
Chamber
wall
Aperture
scattering
Conventional SEM,
pathological scattering
“In-hole Spectrum”
Green =
Extent of
specimen
X-ray sources
NOT excluded
by collimator
Multi-material Faraday Cup “in-hole” test specimen
. Fig. 21.13 Measurement of
“in-hole” spectrum with a
multi-material Faraday cup
consisting of three materials
arranged concentrically
Chapter 21 · Trace Analysis by SEM/EDS