36622
Au, while at E 0 = 2.5 keV, the range for Na collapses to 0.12 μm
(120 nm) in a C matrix to 0.02 μm (20 nm) in Au.22.2.2 Reduced Matrix Absorption
Correction
When the range of X-ray production is reduced by lowering
the beam energy, the generated X-rays undergo lower absorp-
tion because of the reduced path length to the surface. This
can be a strong effect, because X-ray absorption follows an
exponential relationship:I/Ie 0 =−xp ()μρ/ ρs
(22.6)where I 0 is the original intensity and I is the intensity
remaining after passing through a distance s (cm) of a mate-
rial of density ρ (g/cm^3 ) and of mass absorption coefficient,
μ/ρ (cm^2 /g). For strongly absorbed photons, which istypically the case for low energy photons, the matrix correc-
tion for absorption diminishes rapidly (i.e., approaches
unity) as the beam energy is reduced, as shown in. Fig. 22.10
for O K-L2,3 and Cu L 3 -M4,5 in Cu 2 O (measured relative to
MgO and Cu).22.2.3 Accurate Analysis of Low Atomic
Number Elements at Low Beam
Energy
Low atomic number elements with Z ≤ 10 have characteristic
X-ray energies below 1 keV, and these low energy photons
suffer especially high absorption. By minimizing absorption
through operation at low beam energy, Li and Be can be
detected; and B, C, N, and F, can be quantitatively analyzed
with accuracy such that the analyzed value is generally
within ± 5 % relative to the true value, as presented in. Table 22.1 (borides),. Table 22.2 (carbides),. Table 22.3
01.11.00.90.8Absorption correction, A0.70.60.5
246O K-L2,3
Cu L 3 -M4,58
Beam energy (keV)X-ray absorption in Cu 2 O10 12 14 16 18 20. Fig. 22.10 Absorption correc-
tion factor for O K-L2,3 (relative to
MgO) and Cu L 3 -M4,5 (relative to
Cu) as a function of beam energy
. Table 22.1 Analysis of metal borides at E 0 = 5 keV (5 replicates); atomic concentrations
Compound Metal, Cav Relative
accuracy,%σrel, % Boron, Cav Relative
accuracy,%σrel, %CrB 2 0.3482 4.5 0.32 0.6518 −2.2 0.17
CrB 0.5149 3.0 0.17 0.4851 −3.0 0.19
Cr 2 B 0.6769 1.5 0.65 0.3231 −3.1 1.4
TiB 2 0.3373 1.2 3.6 0.6627 −0.6 1.8Chapter 22 · Low Beam Energy X-Ray Microanalysis