Scanning Electron Microscopy and X-Ray Microanalysis

(coco) #1
382

23


There are two “zero-th level” assumptions that underpin the
basis for quantitative electron-excited X-ray microanalysis:


  1. The only reason that the measured X-ray intensity differs
    between the unknown and the standard(s) is that the
    composition is different. No other factors such as the
    specimen shape, orientation, or size influence the
    measured X-ray spectrum.

  2. The specimen is homogeneous in composition over the
    volume excited by the electron beam from which the
    characteristic and continuum X-rays are emitted,
    including the secondary radiation induced by absorption
    of the primary electron-excited radiation.


When either of these conditions is not met, a significant
increase in the overall uncertainty budget of the analysis can
occur beyond the ideal situation in which the uncertainties
arise from counting statistics and from uncertainties in the
calculated matrix correction factors.
Considering “zeroth-level” assumption 1, sample geom-
etry can significantly modify the measured X-ray intensity.
The ideal specimen is flat and placed at known angles to the
incident beam and the X-ray detector(s). Topographic fea-
tures on bulk specimens (defined as those for which the
thickness is much greater than the electron range) or
unusual geometric shapes, such as particles with dimensions
similar to the electron range, can strongly affect the mea-
sured X-rays by modifying X-ray generation and by affect-
ing the loss of X-rays due to absorption. In severe cases, the
impact of “geometric factors” on the final concentrations
becomes so large as to render the compositional results, as
calculated with the conventional standards-based/matrix
corrections protocol or the standardless protocol, nearly
worthless.

23.1 The Origins of “Geometric Effects”:


Bulk Specimens


The ideal sample is compositionally homogeneous on a
microscopic scale, has a flat surface, and is set at known
angles to the incident electron beam and the X-ray spectrom-
eter. Compared to the X-ray spectrum measured from this
ideal spectrum, geometric effects occur when the size and
shape of the specimen (1) modify the interaction of the elec-
trons with the specimen so as to affect the generated X-ray
intensity and (2) alter the length of the absorption path along
which the generated X-rays travel to escape the specimen and
reach the detector so as to affect the measured X-ray
intensity.
Because electron backscattering depends on the local sur-
face inclination to the incident beam, tilted samples generate
fewer X-rays compared to a sample at normal beam inci-
dence (0° tilt). An illustration of this effect for bulk copper, as
calculated with the Monte Carlo simulation embedded in
NIST DTSA-II, is shown in. Fig. 23.1. Even at normal beam
incidence where the backscattered electron (BSE) coefficient

is at a minimum, BSEs carry off energy which would have
gone to cause additional inner shell ionization events fol-
lowed by subsequent X-ray emission had those electrons
remained in the specimen. As the local surface inclination
(tilt) increases, backscattering increases and more X-ray gen-
eration is lost compared to normal beam incidence situation.

. Figure 23.2 shows Monte Carlo calculations of the Cu
K-L2,3 X-ray intensity emitted from a flat, bulk copper speci-
men, expressed as a “k-ratio,” where the denominator is the
intensity emitted from copper at zero tilt (normal beam inci-
dence). As the local surface inclination (tilt angle) increases
above zero degrees, the X-ray production decreases with


0

0.8

0.7

0.6

0.5

0.4
Backscatter coefficien

t

0.3

0.2
20 40
Tilt angle (degrees)

60 80

Cu E 0 = 20 keV

. Fig. 23.1 Backscatter coefficient η vs. surface tilt θ (inclination) for
Cu at E 0 = 20 keV as calculated with the Monte Carlo electron trajectory
simulation embedded in NIST DTSA-II


1.0

0.8

0.6

0.4

k-ratio

0.2

0.0
02040
Tilt angle (degrees)

60 80

Cu K-L2,3 emission vs. Tilt (E 0 = 20 keV)

. Fig. 23.2 Emitted Cu K-L2,3 X-ray intensity calculated as the k-ratio
relative to the intensity at a tilt of 0°, vs. surface tilt θ (inclination), for
Cu at E 0 = 20 keV as calculated with the Monte Carlo electron trajectory
simulation embedded in NIST DTSA-II


Chapter 23 · Analysis of Specimens with Special Geometry: Irregular Bulk Objects and Particles
Free download pdf