451 25
- T, the sample chamber temperature (K): The scattering
skirt is reduced by operating at the lowest possible tem-
perature. - L, the gas path length (m). The shorter the gas path
length, the smaller the gas scattering skirt, as shown in
. Fig. 25.4, where the skirt is compared for gas path
lengths of 3 mm, 5 mm, and 10 mm. Note that the gas
path length appears in Eq. (25.1) with a 3/2 power, so
that the skirt radius is more sensitive to this parameter
than the other parameters in Eq. (25.1). A modification
to the vacuum system of the VPSEM that minimizes the
gas path length consists of using a small diameter tube to
extend the high vacuum of the electron column into the
sample chamber.
25.2.1 Workarounds To Solve Practical
Problems
Gas scattering effects can be minimized but not avoided,
and for many VPSEM measurements and in situ experi-
ments, the microscopist/microanalyst may be significantly
constrained in the extent to which any of the parameters in
Eq. (25.1) can actually be changed to reduce the gas scatter-
ing skirt without losing the advantages of VPSEM opera-
tion. The measured EDS spectrum is always compromised,
but by carefully choosing the problems to study, successful
X-ray analysis can still be performed. A useful way to con-
sider the impact of gas scattering is the general concentra-
tion level at which the remote scattering corrupts the
measured spectrum:
Major constituent: concentration C > 0.1 mass fraction
Minor constituent: 0.01 ≤ C ≤ 0.1
Trace constituent: C < 0.01
Depending on the exact nature of the specimen, gas scat-
tering will almost always introduce spectral artifacts at the
trace and minor constituent levels, and in severe cases arti-
facts will appear at the level of apparent major constitu-
ents.
25.2.2 Favorable Sample Characteristics
Given that the LVSEM operating conditions have been
selected to minimize the gas scattering skirt, what specimen
types are most likely to yield useful microanalysis results? If
most of the gas scattering skirt falls on background material
that contains an element or elements that are different from
the elements of the target and of no interest, then by follow-
ing a measurement protocol to identify the extraneous ele-
ments, the measured spectrum can still have value for
identifying the elements within the target area, always recog-
nizing that the target is being excited by the focused beam
and the skirt.
Particle Analysis
Particle samples comprise a broad class of problems related to
the environment, technology, forensics, failure analysis, and
other areas. Particles are very often insulating in nature so that
a conductive coating is required for examination in the con-
ventional high vacuum SEM, and the complex morphologies
VPSEM 100 Pa 10 mm GPL
1.0
0.9
0.8 He
O 2
0.7
0.6
Cumulative elec
tron intensit
y
0.5
0.4
01020
Radial distance from beam center (micrometers)
30 40 50
. Fig. 25.12 Comparison of the
scattering skirt for He and O 2
25.2 · What Can Be Done To Minimize gas Scattering in VPSEM?