AMPK Methods and Protocols

(Rick Simeone) #1

3.6 Equilibration 1. Open the assay template previously configured (seeSubheading
3.3) on the Seahorse Wave software.



  1. Start the running of the assay by clicking on the run assay
    button on the screen.

  2. Place the sensor cartridge with the utility plate on the machine
    tray taking care of the cartridge orientation (the notch needs to
    be placed on the bottom left).

  3. Confirm the presence of the cartridge to the software by click-
    ing the confirmation button displayed on the screen.

  4. Let the calibration and equilibration step to run.


3.7 Cells Washing
and Preincubation
in Assay Medium



  1. Take the cell plate out of the incubator and check the state of
    the cells under a microscope to verify cell viability, morphology,
    and confluence.

  2. Remove the medium from the each well leaving around 50μlof
    medium to avoid cells from drying during the washing steps.

  3. Add 650μl of pre-warmed MGA or MRA at 37C to each well
    (for glycolysis or mitochondrial respiration assay, respectively).

  4. Repeatsteps 2and 3 two more times.

  5. Replace entirely the medium with 500μl of pre-warmed MGA
    or MRA at 37C well by well (for glycolysis or mitochondrial
    respiration assay, respectively) (seeNote 11).

  6. Ensure cell viability and confluence under the microscope
    (make sure cells did not detach) (seeNote 12).


Incubate the cell plate at 37C in an incubator without CO 2
for 40 min. This incubation time is required for the equilibration of
the medium with the external environment and the stabilization of
cell metabolism after the washing steps (seeNotes 13and 14 ).

3.8 Plate Reading 1. Once the calibration is over, remove the cell plate from the
incubator and confirm the opening of the tray by clicking on
the button displayed on the screen.



  1. Substitute the utility plate with the cell plate on the tray.
    Confirm the presence of the cell plate by clicking on the button
    displayed on the screen. The assay reading should start
    automatically.

  2. At the end of the reading, confirm the discard of the cartridge
    and cell plate on the screen. Empty the tray and discard the
    cartridge in the appropriate waste disposal. Take out the cell
    plate, check the presence of the cells, remove the media, and
    store the cell plate at 20 C if further analyses are needed (see
    Note 15).


298 Claudia Marinangeli et al.

Free download pdf