Encyclopedia of Environmental Science and Engineering, Volume I and II

(Ben Green) #1

1108 STACK SAMPLING


2) The isopropanol bubbler or impinger is eliminated.
An empty bubbler is used for the collection of
liquid droplets.
3) The stack gases are extracted from the sampling
point intermittently over the specified period, usu-
ally 24 hours. An industrial timer is used to cycle
the pump on for at least 12 periods of at least two
minutes each. Between 25 and 60 liters of gas
must be collected.

The EPA Test Method 6B write-up contains detailed
instructions, along with a list of references. It should be
read in detail before the Method is attempted. Testing
should be performed only by personnel trained and expe-
rienced with the equipment and titrations specified in this
method. The Method 6 audit samples are also appropriate
for this method.

Test Method 6C

Test Method 6C is an instrumental method for the determi-
nation of sulfur dioxide from stationary sources. No specific
instrument is required by this method. Ultraviolet, nondis-
persive infrared, or fluorescence instruments can be used
providing they meet the performance specifications and the
test procedures are followed.
The following Measurement System Performance
Specification must be passed by the instrument before actual
environmental samples are analyzed:

1) Analyzer Calibration Error must be less than 
2% of the span for the zero, mid-range, and high-
range calibration gases.
2) Sampling System Bias must be less than  5% of
the span for the zero, mid-range, and high-range
calibration gases.
3) Zero Drift must be less than  3% of the span
over the period of each run.
4) Calibration Drift must be less than  3% of the
span over the period of each run.

The analytical range must be selected such that the SO 2
emission limit required of the source is not less than 30% of
the instrument span. Any run in which the SO 2 concentration
in the stack gas goes off-scale must be repeated.
The EPA Test Method 6C write-up contains detailed
descriptions of the calibration gases required, calibration
procedures, sampling procedures in addition to a list of
references. It should be read in detail before the Method
is attempted. The manufacturer’s instructions will provide
instrument specific instructions. Testing should be per-
formed only by personnel trained and experienced with the
equipment being used.

Test Method 7

Test Method 7 is used for the determination of nitrogen oxide
(NO x ) emissions from stationary sources. In this method the

NO x concentration of a grab sample, collected in an evacu-
ated two liter flask, is measured colorimetrically using the
phenoldisulfonic acid procedure.
The apparatus for sample collection consists of probe,
squeeze bulb, vacuum pump, manometer, and a two liter
flask. The probe is heated if necessary to prevent conden-
sation. A one way squeeze bulb is used to purge the probe
before sampling. The vacuum pump is used to evacuate the
flask. A manometer is used to measure the flask pressure.
The two liter flask contains 25 ml of a H 2 SO 4 /H 2 O 2 solution.
This solution will absorb the NO 2. Oxygen from the source
is required for the oxidation of the NO. If less than 1.5%
oxygen is present in the stack gases, the steps described in
the method to introduce additional oxygen must be used.
Once the flask is evacuated 75 mmHg, valves are rotated to
isolate the flask from the pump, and a sample is introduced
into the flask. It should take less than 15 seconds for the
flask to reach ambient pressure. If a longer time is required,
check the probe for plugging.
After sampling, the flask should sit for at least 16 hours
for the NO to oxidize. The flask is then shaken for two min-
utes before checking the pressure of the flask. The liquid
is then transferred to a container, and the flask rinsed with
distilled water. The pH of the liquid is adjusted to between 9
and 12 with 1 normal NaOH. Just prior to analysis the liquid
is transferred to a 50 ml volumetric flask and diluted to
50 ml. A 25 ml aliquot is transferred to an evaporating dish
and evaporated to dryness. Two ml of the phenoldisulfonic
acid solution is added to the evaporating dish along with
1 ml of distilled water and four drops of concentrated sulfu-
ric acid. The solution is again heated for three minutes fol-
lowed by the addition of 20 ml of water. The solution is then
adjusted to pH 10 with ammonium hydroxide. The solution
is filtered if necessary and transferred to a 100 ml volumetric
flask and diluted to 100ml. The absorbance is measured at
the optimum wave length and compared against standards
for quantification. Audit samples are available through the
EPA’s Emissions, Monitoring and Analysis Division.
The EPA Test Method 7 write-up contains detailed
instructions, along with a list of references. It should be read
in detail before the Method is attempted. As with all of these
methods, testing should be performed only by personnel
trained and experienced with the equipment and the spectro-
photometer used in this method.

Test Method 7A

Test Method 7A is man alternative to Method 7 for the deter-
mination of nitrogen oxide (NO x ) emissions from station-
ary sources. The sample is collected with the same sampling
train used in Method 7. However, instead of using the colori-
metric phenoldisulfonic acid procedure, ion chromatography
is used for quantification. Audit samples, available through
the EPA’s Quality Assurance Division, are required for this
method.
The EPA Test Method 7A write-up contains detailed
instructions, along with a list of references. It should be read
in detail before the Method is attempted. As with all of these

C019_003_r03.indd 1108C019_003_r03.indd 1108 11/18/2005 11:07:16 AM11/18/2005 11:07:16 AM

Free download pdf