1246 WATER AND WASTE MANAGEMENT SYSTEMS IN SPACE
(a potassium monopersulfate compound) combined with sul-
furic acid, or hypochlorite (bleach) before distillation.
Thermoelectric Integrated Membrane
Subsystem (TIMES)
This was the original baseline water recovery subsystem for
the space station (Figure 2). Wastewater is heated to 66C
in a heat exchanger, and is then pumped through bundles of
small diameter hollow Nafion fiber membranes in the evap-
orator module. The Nafion allows only water, gases, and
small neutral molecules to pass through. The pressure on the
exterior of the membrane is reduced to 17 kPa (2.5 psi) to
assist in both transport through the membrane, and subse-
quent evaporation. The water vapor is then condensed, and
the latent heat of condensation is conducted to the input heat
exchanger. The TIMES system was found to produce poorer
quality water than the VCD. 4,5
Vapor Compression Distillation (VCD)
The VCD process involves spreading a thin film of the waste-
water on the inside wall of a thin-walled rotating drum under
low pressure, typically about 4.8 kPa (0.7 psi). The VCD is
shown in Figure 3, and in-depth discussions may be found
in Refs. 4, and 6–11. The system operates at 35C which is
slightly above ambient temperature. Heat is applied to the
outside of the wall causing the thin film of water to boil. The
vapor is extracted from the drum interior and compressed
by a pump. The compressed vapor is then condensed on the
exterior wall of the drum. The compressed vapor condenses
at a higher temperature than that at which it had originally
evaporated. The latent heat of condensation thus supplied
the heat required to evaporate the original feed water in the
inside of the drum. The unevaporated brine, heavily loaded
with contaminants, is recycled back to the inflow stream, or
passed to another subsystem.
TABLE 2 (continued)
QUALITY PARAMETERS Shuttle Potable EPA Current SSF Potable SSF Hygiene
MICROBIAL None Viable — — —
BACTERIA (CFU/100 mL) ————
TOTAL COUNT — — 1 1
ANAEROBES — — 1 1
COLIFORM — 1 1 1
ENTERIC — — — —
VIRUS (PFU/100 mL) — — 1 1
YEAST and MOLD
(CFU/100 mL)
—— 11
RADIOACTIVE CONSTITUENTS
(pCi/L)
NRC LIMITS (see Note3) ——
ORGANIC PARAMETERS (mg/L) (See
Note 2)
————
TOTAL ACIDS — — 500 500
CYANIDE
(total including organic cyanides)
— — 200 200
HALOGENATED HYDROCARBONS — 0.1 (THM)* 10 10
TOTAL PHENOLS — — 1 1
TOTAL ALCOHOLS — — 500 500
TOTAL ORGANIC CARBON (TOC) Reference only — 500 10000
UNCHARACTERIZED TOC (UTOC) — — — —
(see Note 4) — — 100 1000
ORGANIC CONSTITUENTS (mg/L) — — — —
(see Notes 2 and 5) — — — —
Note 1: No detectable gas using a volumetric gas vs. fluid measurement system. This excludes CO 2 used for aesthetics
purposes.
Note 2: MCLs considered independently of others.
Note 3: The maximum contaminant levels for radioactive constituents in potable and personal hygiene water shall
conform to Nuclear Regulatory Commission (NRC) regulations (10CFR20, et al.). These maximum contaminant levels
are listed in the “Federal Register Vol 51, No: 6, 1986, Appendix B, Table 2.
Note 4: Total organic carbon minus identifiable organic contaminants.
Note 5: MCLs for others, if found, will be established as necessary.
* THM = Trihalomethanes
SSF = Space Station Freedom
Source: Ref 3.
C023_001_r03.indd 1246C023_001_r03.indd 1246 11/18/2005 1:29:11 PM11/18/2005 1:29:11 PM