INSTRUMENTATION: WATER AND WASTEWATER ANALYSIS 557
(c) Atomic absorption instrumentation^29
A schematic for a single beam atomic absorption
spectrophotometer is given in Figure 12. Both flame and
electrothermal atomizers may be utilized in this instrumenta-
tion (see Figure 4A). The hollow cathode lamp is the energy
source generating uv or vis radiation that passes through
the sample. The flame or thermal area of the electrothermal
device acts as the sample cell where the sample solution that
has been nebulized (formed into a fine aerosol) is atomized.
(Atomization is the formation of free atoms through thermal
energy in the flame or thermal area.) The flame is generated
from various fuel mixtures: acetylene and the oxidants air,
oxygen or nitrous oxide; hydrogen and the aforementioned
oxidants; and natural gas and air or oxygen. Each of these
mixtures as well as the fuel/oxidant ratio determines the
flame temperature, a critical condition for atomization.
(7) Emission instrumentation 30,31
Emission of emr by elements and some chemical enti-
ties, energized by flames, plasmas, and arc, is the basis of
emission spectroscopy. The electrons are energized and
move to higher energy levels on absorption of the energy.
On relaxation, the electron returns to a lower energy level
and the absorbed energy is emitted as radiation.
Emission methods give rise to atomic spectra by a series
of atomization techniques: namely, flame, inductively cou-
pled argon plasma (ICP), electric arc and spark, and direct
current argon plasma, DCP (see Figure 4B). An emission
spectrophotometer capable of using plasma and arc and
spark sources is illustrated in Figure 13.
(8) Photoluminescence instrumentation
The occurrence of fluorescence and phosphorescence
(photoluminescence) refer to substances which on excitation
by radiation emit light on relaxation of the excited species.
In resonance fluorescence the wavelengths of excitation and
emission are the same. However, in many cases the wave-
length of emitted radiation is longer than that of the exciting
radiation. The difference in fluorescence and phosphores-
cence is the time delay between excitation and emission. The
former is quite small ( 10 ^6 sec), while the latter has a time
delay of several seconds or longer. Fluorescence can occur in
a number of organic and metal-organic complex molecules,
and gases on excitation with uv light.
(a) Uv/visible^32
A general schematic for this instrumentation is given in
Figure 4C. Fluorescence of molecular substances is measured
Field lens
Entrance slit Objective lens
Grating
Wavelength
Light control cam
Exit
slit
Occluder
Sample
Filter
Lamp
Measuring
phototube
FIGURE 10 An example of a simple spectrophotometer. The SPECTRONICR 20. (Courtesy of the Milton Roy Co.,
Rochester, NY)
Reference
Beam
Sample
Beam
Source
M5 Comb
M6
M8
M7
M9
G2
G1
M12
M13
S2
S1
Filters
Thermocouple
M14
C
+
Sampling
Area
Attenuator
M2
M1
M
M
M11
M10
FIGURE 11 Schematic diagram of a typical double-beam infrared spectrophotometer. The sym-
bol M1 through M14 indicate mirrors; S1 and S2 indicate slits; and G1 and G2 indicate gratings.
(Courtesy of the Perkin-Elmer Corporation.)
C009_005_r03.indd 557C009_005_r03.indd 557 11/23/2005 11:12:22 AM11/23/2005 11:12:22