IMPLANTED CARDIAC DEFIBRILLATORS (ICD)

ICDs are placed in patients at high risk for sudden cardiac death from VT
and VFib. They are programmed to analyze the cardiac rhythm, perform anti-
tachycardia pacing, generate an electrical charge, and deliver an electrical
shock. All new generation ICDs have ventricular pacing abilities.

EVALUATION OF ADELIVEREDSHOCK


Patient-reported shocks may be appropriate (underlying VT or VFib), inappro-
priate, or phantom (patients think they got shocked but didn’t). Only interroga-
tion of the device or monitoring during an event can differentiate appropriate
from inappropriate shocks.

Appropriate shocks may result from associated:
■ Electrolyte abnormalities
■ Myocardial ischemia
■ Medications (proarrythmic drugs or med noncomplicance)

Inappropriate shocks may be a result of:
■ Nonsustained VT
■ SVT inappropriately sensed as VT
■ Oversensing T waves as QRS complexes
■ External interference sensed as impulses
■ Broken/displaced ventricular lead

DIAGNOSIS
■ Interrogation of the device
■ External cardiac monitoring during ICD event
■ ECG and labs to evaluate for ischemia or electrolyte abnormalities
■ CXR to evaluate leads
■ All patients with repetitive shock warrant diagnostic evaluation.
■ Stable patient with isolated shock may warrant less extensive evaluation.

TREATMENT
■ Identify and treat contributing conditions.
■ Inappropriate shocks may be prevented with magnet deactivation of the ICD.
■ Second-generation ICDs
■ Deactivate by placing ring magnet over generator for 30 seconds.
■ Reactivate by placing magnet over generator for another 30 seconds.
■ Third-generation ICDs
■ Presence of magnet deactivates ICD (removal reactivates it).
■ Reprogramming, as indicated, by cardiologist

IMPLANATEDCARDIACDEFIBRILLATORFAILURE


CAUSES
May be due to the following:
■ Component failure
■ Lead fracture/displacement
■ Battery depletion
■ Interference with pacemakers

RESUSCITATION