Quadrupole rods
Ion that is unsuccessful
at reaching detector
Detector
RF supply
DC supply
+
+
Ions
Fig. 9.5Quadrupole analyser. The fixed (DC) and oscillating (RF) fields cause the ions to undergo complicated
trajectories through the quadrupole filter. For a given set of fields, only certain trajectories are stable, which
only allows ions of specificm/zto travel through to the detector. The efficiency of the quadrupole is
impaired after a build-up of ions that do not reach the detector. Therefore a set of pre-filters is added to the
quadrupole to remove the ions that would otherwise affect the main quadrupole.
100
Relative ion intensity (%)
900
0
1000 1100
m/z
1200 1300 1400 1500
%
A100
1027.6
A106
988.0
A106
969.4
A97
1059.4A94
1093.1
A86
1154.6
A86
1194.6
A80
1284.0
A78
1317.3
A76
1352.1
M= 102658.3 ±7.0
Fig. 9.6Large intact protein mass accurately measured in electrospray MS. The species of ions are
annotated by the charge state, e.g. with 99, 100, 101 charges, etc., and the associatedm/zvalue. The inset
shows the ‘deconvoluted spectrum’.
361 9.3 Mass analysers