6.5. Photodetectors 399
Example:
If the temperature of a PMT based detector working at 200Kis elevated
to the room temperature of 300K, what would be the relative change in its
Johnson noise.
Solution:
Since all other parameters are supposed to remain constant at the two tem-
peratures, we can write equation 6.5.30 as
σamp = K
√
T
where K =
√
4 FampkBB
Reqv
.
The percentage relative change in Johnson noise can be obtained by writing the
above equation at the two temperatures and taking their relative difference.
σamp =
σamp, 1 −σamp, 2
σamp, 1
× 100
=
√
T 1 −
√
T 2
√
T 1
× 100
=
√
200 −
√
300
√
200
× 100
=22.5%. (6.5.32)
All of the noise components described above contribute to the overall signal-to-
noise ratio. If we know all the individual currents, we can compute the signal-to-noise
ratio from
S/N =
Iγ
[
σ^2 st+σ^2 bk+σ^2 d+σamp^2
] 1 / 2
=
Iγ
[2eμ^2 FB{Ipe+Ibg+Id}+4FampkBTB/Reqv]^1 /^2
. (6.5.33)
However, practically it is difficult to separate the individual currents from the mea-
surements. The easiest and most practical thing to do is to take two sets of mea-
surements, one without and the other with incident light. The shot noise associated
with the measurement without incident light is given by
σbg,d,amp=
√
σ^2 bk+σ^2 d+σ^2 amp, (6.5.34)
while the total shot noise after shining the photocathode with incident light is
σtot=
√
σ^2 st+σ^2 bk+σ^2 d+σ^2 amp. (6.5.35)
Now, the currents corresponding to these noise levels can be subtracted to get the
current due to light only, that is
Ipe=Itot−Ibg,d,amp. (6.5.36)