648 Chapter 11. Dosimetry and Radiation Protection
(^0200400)
++++++ +
+++
++
+++
++
+
(b)
(^0200400)
+
+
+
+
+ ++++
+
+ +
+
+
−
− −
−
−
−−−−−−−−−−
−−−−− − −−−−−− −−−
−− −−
Fixed
Electrode
−−
−−−−−−−−−− −−−−−−
−−− −− − −− −−−− −−
Quartz Fiber
Radiation
(a)
Ionization Chamber
Ionization Chamber
Reading Pointer
Figure 11.4.5: (a) Quartz fiber electroscope without any absorbed
dose. (b) Incident radiation produces charge pairs in the ionization
chamber that decreases the charge on the electrometer pushing the
electrodes close together. The eyepiece of the microscope is shown to
read a dose of 140 units.
Easy Handling:This device is of the size of a pen and can be carried in a
pocket. Its outer metallic surface makes it sturdy against mechanical shocks
and therefore it does not require any special handling requirements.
Quartz fiber electroscope also suffers from some disadvantages, some of which are
listed below.
Readout Errors: The readout scale of the electrometer is fairly narrow and
can not be very accurately read out. The readout process is also prone to
human errors.
Low Dynamic Range:The very small volume of the ionization chamber does
not allow accumulation of large doses and, in hostile radiation environments,
the instrument gets saturated fairly quickly.
Low Accuracy:The accuracy of the instrument is lower than other types of
dosimeters, which does not make it suitable for accurate dose measurements.