684 Chapter 12. Radiation Spectroscopy
trum. A resolution of better than 10−^4 is generally required but is difficult to
achieve. The main reason is the conflicting requirements of the resolution and
the attenuation. For higher resolution the detector should absorb large number
of photons. Recall that the physical limit on resolution is set by the statistical
noise of photon absorption. The larger the number of photons absorbed, the
lower the relative uncertainty ornoise. Since the detector after the sample does
not have the attenuation constraint, it can absorb as many photons as required
to achieve good signal to noise ratio. In most XAFS applications, use to ion-
ization chambers as transmission detectors is fairly common. Such a detector
is based on parallel plate geometry and generally has several chambers.
νi
νf<νi
νi νf<νi
(a)
(b)
X−ray Photon
M
L
K
Fluorescence Photon
Continuum Photoelectron Auger Electron
Hole
X−ray Photon Fluorescence Photon
Continuum
M
L
K
Hole
Photoelectron
Figure 12.1.11: Generation of
(a) fluorescence photon and (b)
Auger electron following x-ray
photon absorption.
Even though it is the most widely used one, transmission measurement is not
the only technique available for absorption spectroscopy. The other two techniques