668 M. Dubernet
obtained is thus is two light beams of the same wavelengths with a phase inter-
ference ofp. According to the phase difference recombination of the interference
signals is constructive or destructive. In other words, the intensity of the interference
signal will vary according to the value ofp. The plot of the variation in intensity as
a function of the phase is called an interferogram and its mathematical model is an
integral. Fourier transform is a mathematical procedure which converts the inter-
ferogram into a signal the intensity of which varies according to wavelength and
which in turn enables the infra-red spectrum to be constructed. This is an intricate
calculation which can only be carried out using powerful computing techniques.
The spectrum thus obtained enables the technique to be used as a classical
method of spectrophotometric analysis, in which specific absorption wavelengths of
organic compounds are employed to determine concentrations of the target analytes.
12.5.3 Installation and Acquisition of Infra-Red Spectra
The sample to be analysed does not require any particular sample pre-treatment.
However, in the case of musts or cloudy wines, a preliminary clarification by cen-
trifugation or filtration is recommended in order to prevent system blockage. Carbon
dioxide levels exceeding 750mg/L should be reduced or eliminated before analysis
to avoid degassing of the sample in the analytical circuit.
The circuit (Fig. 12.2) starts with a sampling needle. A peristaltic pump transfers
the sample into a heating chamber where it is heated to 40◦C. After passing through
a filter, the sample is transferred into the measurement cuvette which is constructed
of fluorosilicate as glass or quartz absorb strongly in the mid infra-red. After analy-
sis, the sample is ejected to waste.
As the complete cycle lasts only 30s, with automation, the analysis rate is of the
order of 120 samples per hour.
12.5.4 Chemometrics
Generally the mid-IR spectrum of a wine or must contains information of analytical
interest the extraction of which in the majority of cases requires highly complex
sampling
needle
Pump
Heating unit
Filter Cuvette
waste
Fig. 12.2Simplified schematic diagram of the analytical circuit