66 Chapter 2
o r
True VU level Instrument indication
Zact
10
600
log
⎛
⎝
⎜⎜
⎜⎜
⎞
⎠
⎟⎟
⎟⎟⎟
where, apparent level instrument indication attenuator or sensitivity indicator.
Thus we can have the following.
- A direct reading from the face of the instrument (zero preferred).
- The reading from the face of the instrument plus the reading from the
attenuator or other sensitivity adjustment—normally a minimum of 4 dB or
higher. When the instrument indicates zero, the apparent level is the attenuator
setting. - The correction factor for impedance other than the reference impedance. 600 Ω is
the normal impedance chosen for a reference, but any value can be used so long
as the voltage across it results in 0.001 W ( Figure 2.16 ).
Example
We have an indication on the instrument of 4 VU. The sensitivity control is at 4. We
are across 50 Ω (a 100-W amplifi er with a 70.7-V output). Using Figure 2.16 , our true VU
would be 4 V U ( 4 VU) 10.8 correction factor 10.8 VU.
Figure 2.15 : Examples of commercial-type VI instrument panels.