Audio Engineering

(Barry) #1

66 Chapter 2


o r


True VU level Instrument indication
Zact

 10


600


log



⎜⎜


⎜⎜




⎟⎟


⎟⎟⎟


where, apparent level  instrument indication  attenuator or sensitivity indicator.


Thus we can have the following.



  1. A direct reading from the face of the instrument (zero preferred).

  2. The reading from the face of the instrument plus the reading from the
    attenuator or other sensitivity adjustment—normally a minimum of  4 dB or
    higher. When the instrument indicates zero, the apparent level is the attenuator
    setting.

  3. The correction factor for impedance other than the reference impedance. 600 Ω is
    the normal impedance chosen for a reference, but any value can be used so long
    as the voltage across it results in 0.001 W ( Figure 2.16 ).


Example
We have an indication on the instrument of  4 VU. The sensitivity control is at  4. We
are across 50 Ω (a 100-W amplifi er with a 70.7-V output). Using Figure 2.16 , our true VU
would be  4 V U  (  4 VU)  10.8 correction factor  10.8 VU.


Figure 2.15 : Examples of commercial-type VI instrument panels.
Free download pdf