100 QUESTIONS IN CARDIOLOGY

(Michael S) #1
form of atrial fibrillation that is commonly associated with rapid

rates at its onset. Occasionally, RF ablation of the AV node is

necessary. Beta adrenergic blockers should be an integral part of

therapy in most ICD patients.

2.Recognise ICD—drug interactions


Antiarrhythmic drugs have the potential for interacting with an

ICD in several ways. Drugs such as flecainide and amiodarone

can increase pacing and defibrillation thresholds. In patients with

a low margin of safety for these parameters, use of these drugs

may result in failure of pacing or defibrillation. Secondly, these

drugs can slow the rate of ventricular tachycardia below the

programmed rate threshold for detection by the ICD; failure of

arrhythmia detection can result. Some rarer interactions include

alteration of the T wave voltage by drugs or hyperkalaemia

resulting in double counting and inappropriate shocks.

3.ICD wound management


As an implanted device, the system is susceptible to infections.

Pain and inflammation of the skin over the ICD may herald an

infective process. Similarly, unexplained fever, particularly

staphylococcal septicaemia may indicate endocarditis involving

the leads and/or tricuspid valve.
Free download pdf