484 CHAPTER 13 Mass Spectrometry and Infrared Spectroscopy
13.1 Mass Spectrometry
At one time, the molecular weight of a compound was determined by its vapor den-
sity or its freezing-point depression, and molecular formulas were determined by
elemental analysis, a technique that determined the relative proportions of the ele-
ments present in the compound. These were long and tedious techniques that re-
quired relatively large amounts of a very pure sample. Today, molecular weights and
molecular formulas can be rapidly determined by mass spectrometry from a very
small amount of a sample.
In mass spectrometry, a small sample of a compound is introduced into an instru-
ment called a mass spectrometer, where it is vaporized and then ionized as a result of
an electron’s being removed from each molecule. Ionization can be accomplished in
several ways. The most common method bombards the vaporized molecules with a
beam of high-energy electrons. The energy of the electron beam can be varied, but a
beam of about 70 electron volts (eV) is commonly used. When the electron beam hits
a molecule, it knocks out an electron, producing a molecular ion, which is a radical
cation—a species with an unpaired electron and a positive charge.
Loss of an electron from a molecule weakens the molecule’s bonds. Therefore,
many of the molecular ions break apart into cations, radicals, neutral molecules, and
other radical cations. Not surprisingly, the bonds most likely to break are the weakest
ones and those that result in the formation of the most stable products. All the
positively charged fragmentsof the molecule pass between two negatively charged
plates, which accelerate the fragments into an analyzer tube (Figure 13.1). Neutral
M
+
+ e−
molecular ion electron
a radical cation
M
molecule
electron
beam
sample
electron
beam
to vacuum pump
magnet
magnet
negatively
charged
accelerating
and focusing
plates
positively
charged
repeller
plate
neutral
molecules
positively charged ions
(deflected according to m/z)
ion exit slit
collector
analyzer tube
recorder
electron
beam
filament
neutral
molecules,
neutral fragments,
and negative
ions
Figure 13.1
Schematic of a mass spectrometer. A beam of high-energy electrons causes molecules
to ionize and fragment. Positively charged fragments pass through the analyzer tube.
Changing the magnetic field strength allows the separation of fragments of varying
mass-to-charge ratio.