Specific Sources of Error with Reagent Strip
Testingcontinued
Urinalysis and Body
Fluids Review 495
INCREASED OR FALSE POSITIVE DECREASED OR FALSE NEGATIVE
Leukocyte esterase
Specific gravity
Sources of error vary with brand of reagent strip. Refer to manufacturer’s package insert.
↑glucose, ↑protein, ↑ascorbic acid, ↑SG; antibiotics;
reading too soon.
Alkaline urine. (Add 0.005 if pH is 6.5 or higher. Correction
is made by automated readers.)
Highly pigmented urine, oxidizing
agents, formalin, nitrofurantoin, vaginal
discharge
↑protein