18012
current. To capture 0.95 of the total beam current requires
a radial distance to approximately 190 μm, as shown in. Fig. 12.7b, and the last 0.05 of the beam electrons are
distributed out to 1000 μm (1 mm). The strong effect of the
gas path length on the skirt radius, which follows a 3/2
exponent in the scattering Eq. 12.1, can be seen in. Fig. 12.7 by comparing the plots for 3-, 5-, and 10-mm
gas path lengths.
1.00.90.80.70.60.50.4
01020
Radial distance from beam center (micrometers)VPSEM 100 Pa O 2Cumulative electron intensity3 mm GPL
5 mm GPL
10 mm GPL30 40 5000.40.50.6Cumulative electron intensity0.70.80.91.03 mm GPL
5 mm GPL
10 mm GPL200
Radial distance from beam center (micrometers)VPSEM 100 Pa O 2400 600 800 1000ab. Fig. 12.7 a Cumulative
electron intensity as a function of
distance (0–50 μm) from the
beam center for 20-keV electrons
passing through 100 Pa of oxygen
as calculated with NIST DTSA-II. b
Cumulative electron intensity as a
function of distance (0–1000 μm)
from the beam center for 20-keV
electrons passing through 100 Pa
of oxygen as calculated with NIST
DTSA-II (GPL = Gas Path Length)
Chapter 12 · Variable Pressure Scanning Electron Microscopy (VPSEM)