Scanning Electron Microscopy and X-Ray Microanalysis

(coco) #1

221 16


position, which you should test to ensure that the positioning
is reproducible between insertions. Manual mechanisms usu-
ally provide a threaded screw with a manual crank to pull the
detector in and out. The threaded screw will usually have a
pair of interlocking nuts which can be positioned to define a
consistent insertion position. The procedure in the sidebar
below will allow you to set and maintain a constant solid
angle and thus also a consistent detector collection efficiency.


z Sidebar: Setting a Constant Detector-to-Sample
Distance to Maintain Solid Angle



  1. Locate the pair of lock nuts on the screw mechanism.
    Move the lock nuts to the inner most position on the
    treaded rod.

  2. Insert the detector as close to the sample as possible.
    Ensure that the detector does not touch the interior
    of the microscope. The detector snout must be
    electrically isolated (no conductive path) from the
    interior of the chamber to eliminate noise caused by
    electrical ground loops.

  3. Twist the upper lock nut to limit the motion of the
    detector towards the sample. Tighten the lower nut
    to lock the upper nut into position.
    4. Test the reproducibility of the insertion point by
    extracting and inserting the detector and collecting a
    series of spectra. If the characteristic peak intensities
    are reproducible (to much better than a fraction of a
    percent) between insertions, the precision is adequate.


The take-off angle is the angle at which X-rays exit a flat sam-
ple in the direction of the detector. For a flat sample mounted
perpendicular to the electron beam at the optimal working
distance, the take-off angle equals the elevation angle. If the
sample is tilted or the sample surface is at a slightly different
working distance, then the take-off angle can be computed
from the sample tilt, the working distance and the sample-to-
detector distance. This is shown in. Fig. 16.13. Often you
will hear the terms elevation angle and take-off angle used
interchangeably. It is more precise however to think of the
elevation angle as being a fixed property of the instrument/
detector geometry and the take-off angle being dependent
upon instrument-specimen configuration.

z Check 1: Verify the Elevation Angle
It is critical that your quantitative analysis software has the
correct elevation/take-off angle. Matrix correction algorithms
use the take-off angle to calculate the correct absorption

Y

X

Z

Axes

Sample

Elevation angle

Sample tilt

Objective
lens

Optimal work

ing distanc
e
Actual work
ing distanc
e

C
L

ψ:

EDS detec

tor snout

. Fig. 16.13 The take-off angle
is a function of the elevation angle,
the sample tilt and the actual
working distance
16. 3 · Practical Aspects of Ensuring EDS Performance for a Quality Measurement Environment

Free download pdf