The manual system, based on our experience, has some advantages
such as it consumes less time in preparation, less expensive, and less
complicated when compared to the other two systems.
For making a tissue microarray block, it is required to follow
the following steps:


  1. Place the “donor” block on a flat table and hold the needle
    perpendicular to the marked position of the “donor” block.

  2. Insert the needle into the “donor” block at the proper depth
    of 5 mm slowly. Do not insert it quickly or too deep to prevent
    from damaging the “donor” block (see Note 3) (Fig. 1b).

  3. Pull out the inner needle so that the tissue is sucked into the
    empty cylinder of the needle (Fig. 1c).

  4. If “donor” block is broken, incubate the easily breaking
    “donor” block in a heating oven or at a room temperature of
    37–40 °C for 15–20 min.

  5. Place the “recipient” block on a table. Hold the needle with
    the extracted tissue perpendicular to the corresponding holes
    of the “recipient” block.

  6. Inject the removed tissue (core) from the “donor” block into
    the corresponding holes of the “recipient” block at the proper
    depth of 4 mm by pushing the plunger slowly (Fig. 1d).

  7. Gently push or tap in all tissue cores to ensure evenness for
    microtome.

  8. Repeat the steps for every sample.


The semi-automated technique is a computer-assisted TMA platform
and designed to provide users for the construction of microarrays and
sample picking for further nucleic acids extraction and protein isola-
tion. The machine could be connected to: the stage controller, the
stage joystick; the digital camera and the personal computer with
software installed (Fig. 2a). The joystick could be used to control the
movement of needles in different directions manually.
The stage controller enables the stage movement either with a
joystick or under PC program. Besides, it takes care of the needle
head vertical and rotational movement. The high-resolution digital
camera is mounted on the machine stand to capture the image of
the blocks. The zooming is manual. The camera can manually be
rotated around its axis. The motorized stage mounting the modu-
lar sample holder is computer assisted. It ensures precise position-
ing of the specimens under the appropriate needle in the operating
conditions. In the standard configuration, up to nine standard his-
tological cassettes can be horizontally mounted in the block holder.
The needles for punching and transfer are easily clamped to their
holder (Fig. 2b). During operation, the selection of the correct
needle is automated and computer controlled. The programmable
vertical displacement allows a precise “donor” core insertion into

3.1 Manual Method
for Making TMA


3.2 Semi-Automated
Method for Making
TMA