Mix 35 ml of solution A with 35 ml of solution B, and make up to 500 ml with
distilled water. Store in a brown bottle at room temperature.
3.Stock urea standard: Prepare 100 mg/dl urea solution in distilled water.
4.Working standard: Prepare 2.5 mg/dl working solution.
18.6 Procedure
- Take a set of nine test tubes and label as blank, standard (S 1 – S 6 ), test for serum
sample (Ts), and test for urine sample (Tu). - Add 0.1, 0.2, 0.4, 0.6, 0.8, and 1.0 ml of working urea standard into the tubes
labeled as S 1 to S 6 , respectively, and make thefinal volume 2 ml with distilled
water. - In Ts add 0.2 ml diluted serum sample (1:10) and 0.4 ml of diluted urine sample
(1:100) to the tube marked as TU, and make the volume of each tube to 2 ml with
distilled water. - Add 2 ml of distilled water in blank tube, and then add 2 ml each of acid reagent
and color reagent to all tubes. - Mix well and incubate all the tubes in boiling water bath for 20C and cool.
- Measure the absorbance at 520 nm in a spectrophotometer.
Reagents Blank S 1 S 2 S 3 S 4 S 5 S 6 TS TU
Working urea
(ml)- 0.1 0.2 0.4 0.6 0.8 1.0 ––
Amount of urea
(mg)- 0.0025 0.005 0.010 0.015 0.020 0.025
Serum/urine
(ml)–––––––0.2 0.4Distilled water
(ml)2 1.9 1.8 1.6 1.4 1.2 1 1.8 1.6Acid reagent
(ml)22 2222222Color reagent
(ml)22 2222222Mix well and keep in boiling water bath for 20 min, and cool and read absorbance at 520 nm16.6 Calculations
Plot a graph between amount of urea at x-axis and absorbance at y-axis. Then
extrapolate amount of urea from the given serum/urine sample from the graph
(include dilution factor for calculation). The urea concentration can also be calcu-
lated by using the following equation:
16.6 Calculations 69