is really measured is the mass-to-charge ratio (m/z) for each ion.) At any given moment,
ions of a particular mass are allowed to pass through the analyser where they are counted
by the detector. Subsequently, ions of a different mass are allowed to pass through the
analyser and again the detector counts the number of ions. In this way, the analyser scans
through a large range of masses.
In the majority of instruments, a particular type of ionisation is coupled to a particular
mass analyser that operates by a particular principle. That is, EI, CI and FAB are combined
with magnetic sector instruments; ESI and its derivatives with quadrupole (or its variant
ion-trap) and MALDI is coupled to TOF detection.
9.3.2 Quadrupole mass spectrometry
The quadrupole analyser consists of four parallel cylindrical rods (Fig. 9.5). A direct
current (DC) voltage and a superimposed radio frequency (RF) voltage are applied to
each rod, creating a continuously varying electric field along the length of the
analyser. Once in this field, ions are accelerated down the analyser towards the
detector. The varying electric field is precisely controlled so that during each stage
of a scan, ions of one particular mass-to-charge ratio pass down the length of the
analyser. Ions with any other mass-to-charge value impact on the quadrupole rods
and are not detected. By changing the electric field (scanning), the ions of different
m/zsuccessively arrive at the detector.
Quadrupoles can routinely analyse up tom/z3000, which is extremely useful for
biological MS since, as we have seen, proteins and other biomolecules normally give
a charge distribution ofm/zthat is centred below this value (see Fig. 9.6).
Note thathexapoleandoctapoledevices are also used, to direct a beam into the
next section of atriple quadrupoleor into the ion trap for example, but the principle
is the same.
Glass capillary ESI needle ±5 kV
Sample solution
Solvent evaporation
Mass
analyser
Heated capillary
Fig. 9.4Electrospray ionisation source. The ESI creates very small droplets of solvent-containing analyte
by atomisation ornebulisationas the sample is introduced into the source through the fine glass (or
other material) hollow needle capillary. The solvent evaporates in the high-vacuum region as the spray of
droplets enters the source. As the result of the strong electric field acting on the surface of the sample
droplets, and electrostatic repulsion, their size decreases and eventually single species of charged analyte
(free of solvent) remain. These may have multiple charges depending on the availability of ionisable
groups.
360 Mass spectrometric techniques