Cardiovascular Fitness Procedures Manual

(coco) #1
„ Confirm with the data manager that the BP monitor is plugged in to an uninterruptible
power supply (UPS). (Enter ‘Confirmed’ with {initials of data manager} in Comments
field).

„ Confirm with the data manager that the treadmill and the UPS for the BP monitor are
on different power circuits. (Enter ‘Confirmed’ with {initials of data manager} in
Comments field).

„ Calibrate SECA scale - Full. If the scale is off by 1 pound or more with the 5 weights
placed on it, inform the MEC manager. A replacement scale will be sent to the MEC.
See Section 2.4.1 for calibration procedures.

„ Calibrate SECA stadiometer. If the reading is off by greater than 1 cm, inform the
MEC manager. Use the stadiometer until a replacement is sent. See Section 2.4.2 for
calibration procedures.

6.4 Data Entry Screens for QC on Equipment


6.4.1 Quality Control Reminder Message Box


Exhibit 6-1. Quality control reminder message box

„ When you log on to the application before the quality control checks are performed,
the system displays a message: ‘One or more of your QC checks have not been
performed.’ See Exhibit 6-1.

„ Click ‘OK’ to this message.