Examples of Equipment/Reagent Quality
Control
Immunohematology Review 484
Blood storage refrigerators & freezers,
platelet incubatorsTemperature recorderAlarmsWaterbathsHeat blocksCentrifugesCell washersPipettesAntiseraReagent cellsAntihuman globulinNot all inclusive. Follow manufacturer’s guidelines & lab’s Standard Operating Procedures Manual.
System for continuous temp monitoring & audible alarm.Compare against thermometer daily. Calibrate as necessary.Check high & low temp of activation quarterly.Check temp daily.Check temp daily. Periodically check each well.Determine optimum speed & time for different procedures upon receipt, after repairs,
& periodically. Check timer every 3 mo, RPM every 6 mo (with tachometer).Check tube fill level daily, AHG volume monthly. Verify time & speed quarterly.Calibrate quarterly.Test with pos & neg controls each day of use. Use heterozygous cells for pos controls.Check for hemolysis. Test each day of use with pos & neg controls.Check anti-IgG activity each day of use by testing Rh-pos cells sensitized with anti-D.