AMPK Methods and Protocols

(Rick Simeone) #1

3.5 Surface Plasmon
Resonance (SPR)


Design a suitable SPR construct for the immobilization of AMPK on
streptavidin chips(seeNote 45).
Followsteps 1– 6 below for capturing AMPK onto streptavidin-based
SPR chips.


  1. Dock the streptavidin sensor chip in the Biacore™3000 instru-
    ment and prime the system 3with HBS-P buffer.

  2. Condition all four flow cells on the chip surface. Set the flow
    rate to 100μl/min and inject 50μl of conditioning solution
    three consecutive times using the Quickinject with Extraclean
    option for the injections.

  3. Normalize all four flow cells on the chip surface: Run the
    normalize procedure found under the Tools tab of the Biacore
    software.

  4. Switch the buffer system to SPR Buffer B by priming 1and
    put instrument in sensorgram mode (found under the Run
    tab).

  5. Dilute the protein to 1.25μg/ml in Buffer B and inject onto
    flow cells 2 and 3 at a rate of 5μl/min in order to achieve
    ~4000–6000 RU of binding to the chip surface.

  6. Flow Buffer B over all four flow cells of the chip surface to
    equilibrate the chip surface at a flow rate of 5μl/min for
    approximately 4 h before collecting compound binding data.

  7. Prepare compound solutions with starting compound concen-
    trations of at least 20-fold over the known EC 50 value while still
    maintaining 2% DMSO in the solutions. The final compound
    solution is then diluted serially in either 3or 2dilutions.
    Use the equationC 1 V 1 ¼C 2 V 2. whereC 1 is the concentration
    of compound stock in DMSO,V 1 volume of DMSO,C 2 start-
    ing concentration of compound for serial dilution, andV 2 final
    volume of compound solution. Solve forC 1 to determine the
    compound stock concentration in DMSO. Example: Com-
    pound X is made up in DMSO stock solutions of 50μM,
    16.7μM, 5.55μM, and 1.85μM. Add 4μl of each compound
    concentration to 196μl of Buffer A, resulting in a final DMSO
    concentration of 2% (v/v) and final compound concentrations
    of 1μM, 333 nM, 111 nM, and 37 nM.

  8. Prepare a DMSO sample curve by adding 5, 10, 20, 30, and
    40 μl of DMSO to 1 ml of assay Buffer A, resulting in final
    concentrations of 0.5, 1, 2, 3, and 4% DMSO.


Followsteps 9– 15 below for acquisition of SPR data of compounds
and their analysis.


  1. Assemble a 96-well sample Greiner Bio-One Plate to include
    the following samples: the DMSO standard curve samples,
    buffer blanks consisting of Buffer B alone, and each


46 Ravi G. Kurumbail et al.

Free download pdf