Instant Notes: Analytical Chemistry

(Tina Meador) #1
the detector response and the mass or concentration of the analyte more
complex curvilinearor logarithmic regressioncalculations are required.

For any analytical procedure, it is important to establish the smallest amount of
an analyte that can be detected and/or measured quantitatively. In statistical
terms, and for instrumental data, this is defined as the smallest amount of an
analyte giving a detector response significantly different from a blank or back-
ground response (i.e. the response from standards containing the same reagents
and having the same overall composition (matrix) as the samples, where this is
known, but containing no analyte). Detection limits are usually based on esti-
mates of the standard deviation of replicate measurements of prepared blanks.
A detection limit of two or three times the estimated standard deviation of the
blanks above their mean, x

_
B, is often quoted, where as many blanks as possible
(at least 5 to 10) have been prepared and measured.
This is somewhat arbitrary, and it is perfectly acceptable to define alternatives
provided that the basis is clear and comparisons are made at the same proba-
bility level.

Where components of a sample other than the analyte(s) (the matrix) interfere
with the instrument response for the analyte, the use of a calibration curve
based on standards of pure analyte may lead to erroneous results. Such matrix
interference effectscan be largely if not entirely avoided by preparing calibra-
tion standards where known amounts of pure analyte are added to a series of
equal sized portions of the sample, a procedure known as spiking. In addition,
one portion of sample is not spiked with analyte. (Note: if spiking sample solu-
tions with analyte changes the volume significantly, volume corrections must be
applied.)
The effects of the matrix on measurements of the analyte in both the spiked
and unspiked samples should be identical. The instrument responses are then
used to construct a calibration graph where the x-axis values are the added
amounts of analyte and the response for the unspiked sample is at x=0 (i.e., the
curve does NOT pass through the origin). The regression line is calculated and
extrapolated back to give a negative intercept on the x-axis at y=0, which corre-
sponds to the amount of analyte in the sample (Fig. 4).
The less reliable procedure of extrapolation rather than interpolation is
outweighed by the advantage of eliminating or minimizing matrix interference.
The method of standard addition is widely used, particularly when the
composition of the sample matrix is variable or unknown so that the response of
a reagent/matrix blank would be unreliable. At least three and preferably more
spiked samples should be prepared, but if the amount of sample is limited, as
few as one sample must suffice. It is especially useful with such analytical tech-
niques as flame and plasma emission spectrometryand potentiometry(Topics
E4, E5 and C8).

Example
The calcium level in a clinical sample was determined by flame emission
spectrometry using a standard addition method, which gave the following data:

Spiked calcium (ppm) 0 10 20 30 40 50
Emission intensity 0.257 0.314 0.364 0.413 0.468 0.528
at 423 nm

Standard
addition


Limit of
detection


46 Section B – Assessment of data

Free download pdf