Detailed calculations of the correlation coefficient, r, and the slope and intercept
values have not been given, but should be set out as in the previous example if a
suitable calculator or computer program is not available.
The amount of calcium in the sample can be read from the extrapolated graph
or calculated from the slope, b, and the intercept, a
Calcium concentration in sample =a/b=0.2569/0.005349 =48 ppm
For some analytical techniques, particularly chromatography, variations in
experimental conditions can adversely affect the precision of the data. A
calibration procedure whereby a constant amount of a selected substance,
the internal standardis added to all samples and analyte standards alike
compensates for variations in sample size and other parameters. The ratio of
the detector response for the analyte in each standard to the corresponding
response for the added internal standard is plotted on the y-axis of a calibration
graph against the mass or concentration of the analyte on the x-axis. The corre-
lation coefficient, slope and intercept values can be computed as shown previ-
ously, and response ratiosfor the analyte and added internal standard in the
samples can then be used to determine the amount of analyte in the samples by
interpolation on the graph. If only one or two analyte standards are prepared,
the amount of analyte in a sample can be calculated by simple proportion, i.e.
=
For some purposes, only the relative amounts of the analytes in a multicomponent
mixture are required. These are normalized to 100 or 1 by expressing each as a
Internal
normalization
sample
standard
for
for
ratio
ratio
response
response
sample
standard
in
in
analyte
analyte
Internal
standardization
B4 – Calibration and linear regression 47
–60 –50 –40 –30 –20 –10 0
0
10 20 30 40 50 60
Sample concentration
48 ppm
Emission intensity at 423 nm
Slope = 0.005349
Spiked samples
Unspiked
sample
Intercept = 0.2569
Calcium (ppm)
0.7
0.6
0.5
0.4
0.3
0.2
0.1
Fig. 4. Standard addition calibration graph, regression line, slope and intercept values for the
flame emission determination of calcium in a clinical sample.